Vital, Vitrea fX and Vitrea Enterprise Suite fX Model number: Vitrea fX 2.1, fX 3.0, and fX 3.1 Vitrea Enterprise Suite (VES) fX 1.2 and VES fX 1.3 The Vitrea 4D CT Brain Perfusion option is intended for post processing based on dynamic CT images continuously acquired during the infection of contrast, for the visualization of apparent blood flow in brain tissue and pictorial illustration of perfusion-related parameters to aid in the assessment of the type and extent of cerebral perfusion disturbances.

Recall: Z-0797-2011 · Initiated October 14, 2010

Recall

Recall Number
Z-0797-2011
Event Number
57219
Firm
Vital Images, Inc.
FEI Number
3001654007
Product Code
LLZ
Status
Terminated
Root Cause
Software design
Initiated
October 14, 2010
Posted
December 21, 2010
Terminated
May 15, 2012
Address
5850 Opus Parkway, Suite 300, Plymouth, MN, 55343-4414

Description

Vital, Vitrea fX and Vitrea Enterprise Suite fX Model number: Vitrea fX 2.1, fX 3.0, and fX 3.1 Vitrea Enterprise Suite (VES) fX 1.2 and VES fX 1.3 The Vitrea 4D CT Brain Perfusion option is intended for post processing based on dynamic CT images continuously acquired during the infection of contrast, for the visualization of apparent blood flow in brain tissue and pictorial illustration of perfusion-related parameters to aid in the assessment of the type and extent of cerebral perfusion disturbances.

Reason

Vital Images Inc., reported that a software defect exists with Vitrea fX 2.1, 3.0 and 3.1 CT Brain Analysis whereby the annotated numeric ROI values on a batched perfusion map are incorrect when created using the auto batching function (from the right click menu) with the auto ROI templates enabled.

Action

Vital Images, Inc sent a Vital "Urgent Software Update Alert" letter dated 14 October 2010 to all Consignees/Customers. The letter described the product and problem. They advised, until new software is installed, that the customers can avoid obtaining incorrect values by tuning off ROI templates prior to selecting the batch option from the right click menu in the CT Brain Analysis application; and recommended that the customers re-examine any ROI perfusion values that were previously generated using automatic collage batching with ROI templates. Please direct any questions or comments to Customer Support at [email protected] or at 1-800-208-3005. Any help needed to install this software update can also be directed to Customer Support.

Distribution

Worldwide distribution: USA including states of: AL, AZ, AR, CA, CO, FL, GA, IA, KY, LA, MD, MA, MI, MN, MO, MT, NY, NJ, ND, OH, PA, PR, TN, TX, WA, WI, and WY, and countries of: CANADA, NETHERLANDS, TAIWAN, CHINA, SINGAPORE, UNITED KINGDOM, BRAZIL, GERMANY, RUSSIA, THAILAND, FRANCE, REPUBLIC OF KOREA, SPAIN, PAKISTAN,IRELAND, JAPAN, AUSTRALIA, INDIA, CZECH REPUBLIC, HONG KONG, EGYPT, ITALY, AUSTRIA, SLOVAK IA, BULGARIA, LEBANON, ALGERA, TAIWAN, SOUTH ARICA, BELGIUM, MALAYSIA, DENMARK, and JORDAN.CANADA, NETHERLANDS, TAIWAN, CHINA, SINGAPORE, UNITED KINGDOM, BRAZIL, GERMANY, RUSSIA, THAILAND, FRANCE, REPUBLIC OF KOREA, SPAIN, PAKISTAN,IRELAND, JAPAN, AUSTRALIA, INDIA, CZECH REPUBLIC, HONG KONG, EGYPT, ITALY, AUSTRIA, SLOVAK IA, BULGARIA, LEBANON, ALGERA, TAIWAN, SOUTH ARICA, BELGIUM, MALAYSIA, DENMARK, JORDAN.

Quantity

182