MEDTRONIC CARELINK 2090 programmer
Recall
- Recall Number
- Z-0795-2019
- Event Number
- 81949
- Firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- FEI Number
- 2182208
- Product Code
- OSR
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- January 17, 2019
- Terminated
- February 2, 2021
- Address
- 8200 Coral Sea St Ne, Mounds View, MN, 55112-4391
Description
MEDTRONIC CARELINK 2090 programmer
There is an error in how the programmer calculates and displays the remaining longevity value during a period of time (up to 2 years) prior to the device reaching its Recommended Replacement Time (RRT).
In the US, beginning 17-Jan-2019, Medtronic Field Representatives hand delivered a courtesy FCA Notification Letter to consignees who, according to Medtronic records, have programmers that are eligible for Micra software updates. Field Representatives will update the identified programmers using a provided USB. Notification will also be hand delivered to Micra Implanting or Follow-up physicians for awareness of the issue.
US, Armenia, Bulgaria, Croatia, Georgia, Hungary, India, Iran, Jordan, Kazakhstan, Latvia, Lebanon, Oman, Pakistan, Qatar, Romania, Saudi Arabia, Syrian Arab Republic
33007 units