FDA Recall Terminated

OPTIGEN Food 30; Catalog number: 85035. For allergen testing. Manufactured by Hitachi Chemical Diagnostics, Mountain View, CA. Distributed by Theridiag, Cedex, France

Recall: Z-0793-2013 · Initiated December 21, 2012

Recall

Recall Number
Z-0793-2013
Event Number
64132
Firm
Hitachi Chemical Diagnostics Inc
FEI Number
2936856
Product Code
DHB
Status
Terminated
Root Cause
Process control
Initiated
December 21, 2012
Posted
February 7, 2013
Terminated
May 16, 2013
Address
630 Clyde Ct, Mountain View, CA, 94043-2239

Description

OPTIGEN Food 30; Catalog number: 85035. For allergen testing. Manufactured by Hitachi Chemical Diagnostics, Mountain View, CA. Distributed by Theridiag, Cedex, France

Reason

There is no evidence that a visual inspection was performed on one tray of coverslips. Possibility of a low fill volume which could lead to a false negative results.

Action

Hitachi sent an Urgent Product Recall Notification letter dated December 21, 2012, to the affected customer. The letter identified the product, the problem, and the action to be taken by the customer. The customer was instructed to discontinue distribution and destroy any remaining inventory of the affected product at their location per their local waste ordinances, determine if any of their customers are using or have inventory of the affected product, rquest that their customers review all patient results reported with the use of the affectted product. The customer was instructed to tell their customers to discontinue use and destroy any remaining inventory per their local waste ordinances. The customer was instructed to retain a copy of the notification letter for their records. The customer was also instructed to send confirmation of the destruction within 5 days Customers were instructed to complete and return the attached Certificate of Destruction. For questions customers were instructed to contact their local area Customer Service. For questions regarding this recall 650-961-5501.

Distribution

Worldwide Distribution - France only

Quantity

20 kits