FDA Recall Terminated

Mizuho OSI Trios, Table Base; Voltage: 100-240V ~ 50/60 Hz Amps: 5,0-5.0A; Model 7803 and 7803AP Orthopaedic: Operating room tablebase - an electronically powered mobile operating table is created, designed for temporary (<24 hours) support and positioning of a patient in a prone, supine, or lateral position depending on the specific table top utilized.

Recall: Z-0792-2016 · Initiated December 14, 2015

Recall

Recall Number
Z-0792-2016
Event Number
72937
Firm
Mizuho OSI
FEI Number
2921578
Product Code
GDC
Status
Terminated
Root Cause
Employee error
Initiated
December 14, 2015
Posted
February 9, 2016
Terminated
February 10, 2016
Address
30031 Ahern Ave, Union City, CA, 94587-1234

Description

Mizuho OSI Trios, Table Base; Voltage: 100-240V ~ 50/60 Hz Amps: 5,0-5.0A; Model 7803 and 7803AP Orthopaedic: Operating room tablebase - an electronically powered mobile operating table is created, designed for temporary (<24 hours) support and positioning of a patient in a prone, supine, or lateral position depending on the specific table top utilized.

Reason

Table Bases may have been assembled incorrectly, missing a retaining ring from the head end assembly. Over time, the parts could go out of alignment, making clutch engagement or disengagement difficult. If the clutch cannot disengage, the user would be unable to rotate the patient from supine to prone.

Action

All consignees were notified by telephone and a recall letter was mailed to each facility as a follow up. Customers were asked to remove tables from operation until they were inspected by OSI service personnel.

Distribution

US, South Africa, Japan and United Arab Emirates.

Quantity

33