FDA Recall Open, Classified

VITEK 2 REAGENT BCL TEST KIT VTK2 20 CARDS, CATALOG 21345

Recall: Z-0790-2023 · Initiated November 9, 2022

Recall

Recall Number
Z-0790-2023
Event Number
91148
Firm
Biomerieux Inc
FEI Number
1950204
Product Code
LQL
Status
Open, Classified
Root Cause
Storage
Initiated
November 9, 2022
Posted
January 4, 2023
Address
595 Anglum Rd, Hazelwood, MO, 63042-2320

Description

VITEK 2 REAGENT BCL TEST KIT VTK2 20 CARDS, CATALOG 21345

Reason

As the temperature and time out of range was exceeded, product performances cannot be guaranteed

Action

Biomerieux sent a FIELD SAFETY CORRECTIVE ACTION notice to the US subsidiary on 11/09/2022 by email. The notice requested the subsidiary identify their customers and issue an URGENT FIELD SAFETY NOTICE which explains the problem and request the following actions: "1. Confirm this letter has been distributed to and reviewed by all appropriate personnel within your organization. 2. Check your inventory for any of the impacted products (reference Table 1) associated with local shipments. 3. Stop using and scrap any remaining impacted products in your inventory. a. Request product replacement or credit for the destroyed product. b. Document the quantity destroyed in the Product Information section of the attached Acknowledgement Form. 4. Please store this letter with your bioMrieux documentation. 5. Complete the attached Acknowledgement Form and return it to your local bioMrieux representative as soon as possible." The customer notices were sent by FedEx Overnight Letter on 18 NOV 2022.

Distribution

US Nationwide distribution.

Quantity

1 unit