FDA Recall Terminated

Superset with 22MM ID, Double Swivel Elbow 22MM OD/15MM ID and Bronchoscopy Port, REF 3503031. Airway connector.

Recall: Z-0786-2020 · Initiated August 7, 2019

Recall

Recall Number
Z-0786-2020
Event Number
83527
Firm
Intersurgical Inc
FEI Number
1000120301
Product Code
BZA
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
August 7, 2019
Terminated
December 15, 2020
Address
6757 Kinne St, East Syracuse, NY, 13057-1215

Description

Superset with 22MM ID, Double Swivel Elbow 22MM OD/15MM ID and Bronchoscopy Port, REF 3503031. Airway connector.

Reason

The port may crack in the swivel elbow where the flip cap is inserted, and these cracks could possibly leak if they reach a significant size.

Action

Intersurgical Inc. issued Urgent Medical Device Recall notices to customers dated August 5, 2019. Customers were advised to take the following actions: 1. Immediately discontinue use and quarantine any stock with the product/lot number described in the recall notification. 2. If you have distributed any of these products, please immediately contact your customers, advise them of the recall situation and have them return their outstanding recalled stock to you. 3. Contact Intersurgical Customer Support at 800-828-9633 for a Return Goods Authorization (RGA) Number. Hours of operation are Monday through Friday, 8:00 a.m. to 4:30 p.m. EST. a. Once you have received the RGA Number, please enter it in the space provided on the Customer Product Recall Acknowledgement Form: i. Return the completed form via email to Intersurgical Inc. at [email protected]. ii. The subject line should contain your Company name and RGA number. 1. This will allow us to document your receipt of this letter and the amount of product you have on-hand for return. b. Intersurgical will credit your account when the product is received. 4. Please note: If you do not have the recalled product in stock, please check the appropriate box on the Customer Product Recall Acknowledgement Form. a. Return the completed form via email to Intersurgical Inc. at [email protected]. b. The subject line should contain your Company name and Intersurgical Recall

Distribution

US Nationwide

Quantity

19,200 pieces