MAC-LAB 5000EX systems
Recall: Z-0785-03
·
Initiated April 14, 2003
Recall
- Recall Number
- Z-0785-03
- Event Number
- 25951
- Firm
- General Electric Medical Systems Information Technology
- FEI Number
- 2124823
- Product Code
- KMI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 14, 2003
- Posted
- April 25, 2003
- Terminated
- June 24, 2004
- Address
- 8200 W. Tower, Milwaukee, WI, 53223
Description
MAC-LAB 5000EX systems
Reason
There is an incorrect jumper set-up on a motherboard of the devices that could eventually cause a polyfuse to overheat resulting in smoke.
Action
A letter was sent to the consignees on April 14, 2003 and requested that the consignees correct the device or contact the firm to have them perform the correction.
Distribution
The recalled products were shipped to hospitals located nationwide in the United States and to other consignees located worldwide.