FDA Recall Terminated

TomoTherapy Treatment System - Product Usage: used as an integrated system for the planning and precise delivery of radiation therapy, stereotactic radiotherapy, or stereotactic radiosurgery to tumors or other targeted tissues while minimizing the delivery of radiation to vital healthy tissue.

Recall: Z-0784-2021 · Initiated December 10, 2020

Recall

Recall Number
Z-0784-2021
Event Number
86993
Firm
Accuray Incorporated
FEI Number
3003873069
Product Code
IYE
Status
Terminated
Root Cause
Software Design Change
Initiated
December 10, 2020
Terminated
September 26, 2022
Address
1240 Deming Way, Madison, WI, 53717-1954

Description

TomoTherapy Treatment System - Product Usage: used as an integrated system for the planning and precise delivery of radiation therapy, stereotactic radiotherapy, or stereotactic radiosurgery to tumors or other targeted tissues while minimizing the delivery of radiation to vital healthy tissue.

Reason

"MLC tickle error" may result in the delivered dose to effectively rotate from the planned dose.

Action

Urgent Medical Device Recall "MLC tickle error" issue during TomoHelical deliver was sent to customers beginning 12/10/2020. Accuray is planning software releases that will address this issue. All TomoTherapy systems at affected software versions will be upgraded to these versions once released.

Distribution

Worlwide distribution - US Nationwide distribution including in the states of CA, FL, IL, ND, NJ, OK, UT and the countries of Australia, China, Italy, Switzerland, United Arab Emirates.

Quantity

7 units