FDA Recall Terminated

Anspach Effort, Inc., 98-0061 MA-15C Bearing Sleeve with Removable Bur Guard. A reusable device used with dissection tools; designed for Transphenoidal and Skull base procedures.

Recall: Z-0783-2013 · Initiated September 5, 2012

Recall

Recall Number
Z-0783-2013
Event Number
63793
Firm
The Anspach Effort, Inc.
FEI Number
1045834
Product Code
HTT
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
September 5, 2012
Posted
February 4, 2013
Terminated
August 13, 2014
Address
4500 Riverside Drive, Palm Beach Gardens, FL, 33410-4235

Description

Anspach Effort, Inc., 98-0061 MA-15C Bearing Sleeve with Removable Bur Guard. A reusable device used with dissection tools; designed for Transphenoidal and Skull base procedures.

Reason

Contact between the bur and bur guard could generate metal fragments that may or may not be visible to the surgeon and can potentially remain in the surgical site. The materials used to fabricate the bur guard are not traceable, design validation did not effectively evaluate adequate protection of adjacent tissue and inspection results for each of these bur guards were not documented.

Action

The firm, Anspach Effort, Inc. sent an "URGENT: Medical Device Removal" letter dated November 7, 2012, to its customer. The letter identified the product, problem and actions to be taken. The customer was instructed to do the following: 1) Screen their inventory and remove and return all products immediately. 2) Complete and return the attached Customer Reply Form via fax or email to the address provided on the form. Should the customer have any questions, please contact Anspach Product Support at (800) 327-6887.

Distribution

Distributed in the state of Massachusetts

Quantity

8 devices