ARCHITECT Folate Reagent Kit, list 6C12-20 ( 4 x 100 tests), 6C12-25 (1 x 100 tests) and 6C12-30 (4 x 500 tests), Abbott Laboratories, Abbott Park, IL 600064 U.S.A.

Recall: Z-0783-05 · Initiated March 11, 2005

Recall

Recall Number
Z-0783-05
Event Number
31410
Firm
Abbott Laboratories MPG
FEI Number
1415939
Product Code
CGN
Status
Terminated
Root Cause
Other
Initiated
March 11, 2005
Posted
April 28, 2005
Terminated
January 18, 2006
Address
100 Abbott Park Rd, Abbott Park, IL, 60064-3502

Description

ARCHITECT Folate Reagent Kit, list 6C12-20 ( 4 x 100 tests), 6C12-25 (1 x 100 tests) and 6C12-30 (4 x 500 tests), Abbott Laboratories, Abbott Park, IL 600064 U.S.A.

Reason

concentrations for normal samples generated with current reagent lots of the ARCHITECT Folate assay are lower than data generated for the package insert.

Action

Abbott international affilates were notified of the recall by e-mail on 03/11/2005, and requested to notifiy their customers. The letters inform the end users of the downward shift in control and/or patient results, and requested them to assess their normal and deficient ranges for their specific population, assess their quality control ranges and follow their laboratory procedures for reviewing patient results and notifying Health Care Providers that they serve. There was no U.S. distribution.

Distribution

Canada, Singapore, New Zealand, Germany, Hong Kong, Japan, and Australia

Quantity

2,521 kits