FDA Recall
Terminated
SOPRO Camera Control Units, PAL, SV420P, REF (Catalog) #72200506 v1.1; Made in France; Manufactured for Smith & Nephew, Inc., Endoscopy, Andover, MA 01810.
Recall: Z-0782-2007
·
Initiated January 19, 2007
Recall
- Recall Number
- Z-0782-2007
- Event Number
- 37819
- Firm
- Smith & Nephew, Inc Endoscopy Division
- FEI Number
- 1643264
- Product Code
- FWF
- Status
- Terminated
- Root Cause
- Other
- Initiated
- January 19, 2007
- Posted
- May 5, 2007
- Terminated
- September 2, 2008
- Address
- 76 S. Meridian Ave, Oklahoma City, OK, 73107-6512
Description
SOPRO Camera Control Units, PAL, SV420P, REF (Catalog) #72200506 v1.1; Made in France; Manufactured for Smith & Nephew, Inc., Endoscopy, Andover, MA 01810.
Reason
Surgical camera control units were not designed and manufactured to pass the IEC 60101 testing for electrical fast transients (EFT)
Action
Recalling firm sent Recall notification letters to customers (Jan 2007) and distribution centers requesting device return.
Distribution
Foreign Distribution Only-No U. S. A. distribution International distribution to: Argentina, Austria, Australia, Belgium, Brazil, China, Columbia, Denmark, Dubai, France, Germany, Greece, Italy, India, Japan, Korea, Malaysia, Netherlands, Norway, Portugal, South Africa, Spain, Sweden, Turkey and United Kingdom.
Quantity
25 units.