FDA Recall Terminated

SOPRO Camera Control Units, NTSC, SV420P; REF (Catalog) #72200505 v1.1; Made in France; Manufactured for Smith & Nephew, Inc., Endoscopy, Andover, MA 01810.

Recall: Z-0781-2007 · Initiated January 19, 2007

Recall

Recall Number
Z-0781-2007
Event Number
37819
Firm
Smith & Nephew, Inc Endoscopy Division
FEI Number
1643264
Product Code
FWF
Status
Terminated
Root Cause
Other
Initiated
January 19, 2007
Posted
May 5, 2007
Terminated
September 2, 2008
Address
76 S. Meridian Ave, Oklahoma City, OK, 73107-6512

Description

SOPRO Camera Control Units, NTSC, SV420P; REF (Catalog) #72200505 v1.1; Made in France; Manufactured for Smith & Nephew, Inc., Endoscopy, Andover, MA 01810.

Reason

Surgical camera control units were not designed and manufactured to pass the IEC 60101 testing for electrical fast transients (EFT)

Action

Recalling firm sent Recall notification letters to customers (Jan 2007) and distribution centers requesting device return.

Distribution

Foreign Distribution Only-No U. S. A. distribution International distribution to: Argentina, Austria, Australia, Belgium, Brazil, China, Columbia, Denmark, Dubai, France, Germany, Greece, Italy, India, Japan, Korea, Malaysia, Netherlands, Norway, Portugal, South Africa, Spain, Sweden, Turkey and United Kingdom.

Quantity

77 units.