FDA Recall Terminated

Halo Support Bracket Assembly (Part #438-B1010). The bracket is sold as a component of the BUDDE Halo Neurosurgical Retractor System, or sold as a separate component. The defective sub-assembly part is the Clamp Assembly (Part #438A1071), which is part of the Halo Bracket Assembly.

Recall: Z-0781-03 · Initiated February 26, 2003

Recall

Recall Number
Z-0781-03
Event Number
25715
Firm
Ohio Medical Instrument Co Inc
FEI Number
1525725
Product Code
GZT
Status
Terminated
Root Cause
Other
Initiated
February 26, 2003
Posted
April 25, 2003
Terminated
December 1, 2003
Address
4900 Charlemar Dr, Cincinnati, OH, 45227

Description

Halo Support Bracket Assembly (Part #438-B1010). The bracket is sold as a component of the BUDDE Halo Neurosurgical Retractor System, or sold as a separate component. The defective sub-assembly part is the Clamp Assembly (Part #438A1071), which is part of the Halo Bracket Assembly.

Reason

The component was manufactured with an insufficient weld that could break during use.

Action

The firm contacted their accounts by email and telephone on 2/26/2003.

Distribution

The units were shipped to domestic consignees in NY, NJ, IL, and FL; and to foreign customers located in Japan, Hong Kong, Korea, Great Britian, and Germany.

Quantity

43 components