MRI Eclipse 1.5T System Model #781230. The 1.5T Magnetic Resonance Imaging system intended for use with clinical/radiologic diagnostic imaging applications.

Recall: Z-0780-2010 · Initiated December 7, 2009

Recall

Recall Number
Z-0780-2010
Event Number
53989
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
1525965
Product Code
LNH
Status
Terminated
Root Cause
Other
Initiated
December 7, 2009
Posted
February 19, 2010
Terminated
September 19, 2012
Address
595 Miner Road, Cleveland, OH, 44143-2131

Description

MRI Eclipse 1.5T System Model #781230. The 1.5T Magnetic Resonance Imaging system intended for use with clinical/radiologic diagnostic imaging applications.

Reason

Polymeric RF connector block on the tabletop overheated and eventually ignited during scanning due to combination of arcing of electrical components associated with the RF trap assembly located below the connector and flame rating of the connector block material.

Action

An "Urgent-Field Safety Notice" dated December 4, 2009 was issued via certified mail to customers. The notification letter described the affected product, problem, hazard involved and action to be taken by customer. The customer will ensure to maintain frequent contact with patients during scanning and that the patient comfort kit is installed for all patient studies. The Customer will be visited by a Philips Service Representative to inspect their MRI System and install the newly remanufactured RF Connector Block. Additionally, step-by-step trap/cable installation instructions including pictures are being drafted to ensure correct installation of cabling. If you need any further information or support concerning this issue, please contact your local Philips representative or 1-800-722-9377.

Distribution

Worldwide distribution- United states: (AL, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, LA, MA, MD, ME, MN, MO, NE, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA & WI), ARGENTINA, BRAZIL, CANADA, CHINA, CYPRUS, FRANCE, GERMANY, GREECE, ITALY, JAPAN, KOREA, NEW ZEALAND, PAKISTAN, POLAND, RUSSIA, SWITZERLAND, TAIWAN & UNITED KINGDOM.

Quantity

290 units