FDA Recall Open, Classified

VERIFY Dual Species Self-Contained Biological Indicators, 50 per box Item Number: S3060 Intended for use in installation testing and routine monitoring of steam and ethylene oxide sterilization processes

Recall: Z-0778-2023 · Initiated December 12, 2022

Recall

Recall Number
Z-0778-2023
Event Number
91315
Firm
Steris Corporation
FEI Number
1527821
Product Code
FRC
Status
Open, Classified
Root Cause
Employee error
Initiated
December 12, 2022
Posted
January 3, 2023
Address
5960 Heisley Rd, Mentor, OH, 44060-1834

Description

VERIFY Dual Species Self-Contained Biological Indicators, 50 per box Item Number: S3060 Intended for use in installation testing and routine monitoring of steam and ethylene oxide sterilization processes

Reason

Biological indicator (BI) used to promote growth of the G. stearothermophilus and B. atrophaeus spores was inconsistent in promoting growth throughout lot , some BIs within the lot performed as intended while others did not. When used to monitor steam and ethylene oxide sterilization cycles, the potential hazard is that the BI will show no growth indicating that the cycle was effective, even if it may not have been.

Action

Steris issued Urgent Medical Device Recall letters to Customers and Distributors on 12/12/2022 via FedEX. Letter states reason for recall, health risk and action to take: 1. Please immediately inspect your on-hand inventory for product affected by this recall. 2. Please complete the Medical Device Recall Response Form included with this letter and destroy any remaining product in your inventory. STERIS will coordinate shipment of replacement product upon receipt of the completed Recall Response Form. 3. Return the completed Recall Response Form via email to: [email protected] or via fax to (440) 392-8963. If you have questions regarding this recall, please contact Melissa Gonsalves, Senior Product Manager at (440) 392-7043, or STERIS Customer Service at 1-800-548-4873.

Distribution

Nationwide Foreign: AE AT CA CO ES FI GU IN IT JP MX SG TH TW ZA

Quantity

338 boxes