FDA Recall Terminated

GEM Premier 4000 Laboratory Analyzer

Recall: Z-0778-2007 · Initiated March 27, 2007

Recall

Recall Number
Z-0778-2007
Event Number
37719
Firm
Instrumentation Laboratory Co.
FEI Number
1217183
Product Code
CHL
Status
Terminated
Root Cause
Other
Initiated
March 27, 2007
Posted
May 3, 2007
Terminated
May 23, 2012
Address
113 Hartwell Ave, Lexington, MA, 02421-3125

Description

GEM Premier 4000 Laboratory Analyzer

Reason

Instrument in Micro Sampling Mode failed to aspirate and no error code was generated and an incorrect patient report could result

Action

Instrumentation Labs on 3/2707 notified Service Reps to perform on site mandatory upgrade. The foreign counterparts were advised by email on 3/21/07

Distribution

Worldwide, including USA, Australia, Austria, Greece, Spain, Netherlands, Great Britain, France, Germany, Italy, and Sweden

Quantity

210