FDA Recall Terminated

Varian brand Clinac 600C (CD, EX) and Portal Vison with Dynamic MLC or RPM system capability

Recall: Z-0777-05 · Initiated February 4, 2005

Recall

Recall Number
Z-0777-05
Event Number
31234
Firm
Varian Medical Systems Inc 3100 Hansen Way M/S E-210 Palo Alto CA 94304-1030
FEI Number
2916710
Product Code
IYE
Status
Terminated
Root Cause
Software design
Initiated
February 4, 2005
Posted
April 27, 2005
Terminated
August 17, 2006

Description

Varian brand Clinac 600C (CD, EX) and Portal Vison with Dynamic MLC or RPM system capability

Reason

Treatment beam may not completely pause while device changes field shape; Potential for mistreatment could occur.

Action

On 2/4/05, the firm issued letters to all its direct consignees, informing them of the affected product and providing instructions on the device correction.

Distribution

The firm has indicated that product was released for distribution to 143 direct consignees in US and to 25 foreign countries via international distribution centers. Foreign distribution include: Australia, Bahrain, Belgium, Canada, Czech Republic, Egypt, Finland, France, Germany, Iceland, India, Israel, Italy, Japan, Kazakhstan, Latvia, Norway, Poland, Portugal, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, and United Kingdom.

Quantity

170 units