FDA Recall Terminated

Intellivue iX Information Center Software - All PIIC iX Surveillance stations including: 866023 IntelliVue Info Center iX A.0 866117 PIIC Classic Upgrade 866389 IntelliVue Info Center iX B.0, C.0 867141 IntelliVue Info Center iX B.0 866025 IntelliVue Server IX A.0 866118 Database Server Upgrade A.0

Recall: Z-0776-2018 · Initiated January 3, 2018

Recall

Recall Number
Z-0776-2018
Event Number
79132
Firm
Philips Electronics North America Corporation
FEI Number
1218950
Product Code
MHX
Status
Terminated
Root Cause
Software design
Initiated
January 3, 2018
Terminated
November 14, 2019
Address
3000 Minuteman Rd, Andover, MA, 01810-1032

Description

Intellivue iX Information Center Software - All PIIC iX Surveillance stations including: 866023 IntelliVue Info Center iX A.0 866117 PIIC Classic Upgrade 866389 IntelliVue Info Center iX B.0, C.0 867141 IntelliVue Info Center iX B.0 866025 IntelliVue Server IX A.0 866118 Database Server Upgrade A.0

Reason

Once a surveillance station is restarted on January 1, 2018 or later, the station will be unable to perform patient discharge and transfer operations. Any subsequent attempt to perform these operations will cause the station to restart, resulting in a short period of loss of monitoring at the Surveillance station during such restart. Until this issue can be corrected, users should avoid intentionally restarting their Surveillance stations in 2018.

Action

To avoid this issue, do not intentionally restart any surveillance stations that have been running normally since the new year. You may use the workflows described in attachment Alternative Workflows for PIIC iX to avoid operations that may cause the surveillance station to restart. When a patient is discharged, confirm that the patient is discharged. Philips has initiated a correction to address this issue and will provide this software update to customers with affected devices at no charge. Software updates for PIIC iX A.0x and B.0x are expected to be available by January 8th. A Philips Healthcare representative will contact customers with affected devices to arrange for correction of the issue. Contact your local Philips representative if you have a question about any device affected by this issue.

Distribution

Domestic and Foreign

Quantity

20357