Trilogy Acetabular AB Shells with Cluster Holes. The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage.
Recall
- Recall Number
- Z-0774-2013
- Event Number
- 63531
- Firm
- Zimmer, Inc.
- FEI Number
- 1000220733
- Product Code
- MRA
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- September 14, 2012
- Posted
- February 4, 2013
- Terminated
- June 20, 2014
- Address
- 345 E Main St, Warsaw, IN, 46580-2746
Description
Trilogy Acetabular AB Shells with Cluster Holes. The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage.
Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy Acetabular System shells.
Zimmer sent an "URGENT MEDICAL DEVICE NOTIFICATION" letter dated September 20, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.The firm expanded the scope of the recall in January 2013 to include devices that use a secondary mean of fixation. New recall letters will be submitted to distributors and surgeons.
Worldwide Distribution-USA (nationwide) and the countries of Denmark, Germany, Greece, Hungary, Ireland, Italy, Netherlands, Spain, Sweden, UK, Poland, France, Australia, Chile, Canada, India, Japan, New Zealand, Peru, South Africa, Colombia, Ukraine, and UAE.
275,952 units