Leica M220 F12 surgical microscope The Leica M220 F12 surgical microscope is an optical instrument for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.
Recall
- Recall Number
- Z-0772-2019
- Event Number
- 81903
- Firm
- Leica Microsystems, Inc.
- FEI Number
- 3000210133
- Product Code
- HRM
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- January 2, 2019
- Terminated
- April 2, 2021
- Address
- 1700 Leider Ln, Buffalo Grove, IL, 60089-6622
Description
Leica M220 F12 surgical microscope The Leica M220 F12 surgical microscope is an optical instrument for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.
The M220 optics may unintentionally drop into the surgical field, risking contact with the patient.
The firm, Leica, disseminated an "URGENT FIELD SAFETY NOTICE" dated 12/18/2018 on 01/02/2019 to its customers. The notice described the product, problem and actions to be taken. The notice directed the user to inspect the device for a gap between the black adapter and the white housing or a free rotation of the optics carrier for more 360o. If either are noted, the device could not be used. If you have any questions, contact the Clinical & Scientific Affairs Specialist at 847-876-5705 or email to: [email protected].
Worldwide distribution: US (nationwide) and countries of: Albania, Algeria, Armenia, Argentina, Bangladesh, Brazil, Burundi, Canada, China, Colombia, Egypt, France, Germany, Guatemala, India, Indonesia, Japan, Nepal, Netherlands, New Zealand, Malaysia, Pakistan, Peru, Philippines, Poland, Qatar, Romania, Russia Federation, Saudi Arabia, Senegal, Spain, South Africa, South Korea, Switzerland, Tanzania, Thailand, Turkey, Ukraine, and Vietnam.
15 units