FDA Recall Terminated

Dimension(R) Creatinine Flex(R) Reagent Cartridges. The CREA method is an in vitro diagnostic test for the quantitative measurement of creatinine in human serum, plasma, and urine on the Dimension Vista System.

Recall: Z-0772-2012 · Initiated November 22, 2011

Recall

Recall Number
Z-0772-2012
Event Number
60618
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
CGX
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
November 22, 2011
Posted
January 15, 2012
Terminated
April 3, 2014
Address
500 GBC Drive PO BOX 6101, Mailstop 514, Newark, DE, 19714-6101

Description

Dimension(R) Creatinine Flex(R) Reagent Cartridges. The CREA method is an in vitro diagnostic test for the quantitative measurement of creatinine in human serum, plasma, and urine on the Dimension Vista System.

Reason

Siemens has confirmed the potential for falsely elevated creatinine results with CREA FLex(R) reagent cartridge lots GA2196, GA2203, DC2221, DB2221, GA2229 and FA2237 when processing tests from a well that has been punctured.

Action

Siemens sent an "URGENT FIELD SAFETY NOTICE" dated November 2011 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter recommends that customers discontinue and discard any remaining inventory of the affected product. The firm also recommends customers to test QC at least every 8 hours with lots manufactured after lot FA2237. Additionally, a Field Correction Effectiveness Check form was attached to the letter for customers to complete and return via fax to 302-631-8467. Contact the Siemens Customer Service Center at 888-588-3916 for questions regarding this notice.

Distribution

Worldwide Distribution-USA (nationwide) including the District of Columbia and the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY and countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Benin, Bosnia Herzegovina, Botswana, Brazil, Canada, Canary Islands, Ceuta/Mililla, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Fiji, France, French Guinea, Germany, Great Britian, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxemburg, Macedonia, Mexico, Montenegro, Morocco, Nepal, Nether Antilles, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Uruguay, Venezuela, Vietnam, Yemen and Zimbabwe.