FDA Recall Terminated

SOMATOM Force, Definition AS, Definition Edge, Definition Flash, Drive, Confidence and Edge Plus

Recall: Z-0771-2021 · Initiated November 24, 2020

Recall

Recall Number
Z-0771-2021
Event Number
87079
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
JAK
Status
Terminated
Root Cause
Software design
Initiated
November 24, 2020
Terminated
February 6, 2024
Address
40 Liberty Blvd, Malvern, PA, 19355

Description

SOMATOM Force, Definition AS, Definition Edge, Definition Flash, Drive, Confidence and Edge Plus

Reason

System does not trigger a cancel command followed by a reload of the scan with the new parameter(s). This issue may result in a delay in diagnosis and/or need for patient rescan

Action

Customer notification letters will be sent starting December 14, 2020 to all affected customers via Customer Advisory Notice CT057/20/S. The technical solution, software versionVB20A_SP3, will be distributed to all affected customers free of charge via CT056/20/S. For questions or technical support, contact your local application specialists or your local service/sales organization at 1-800-888-7436.

Distribution

US Nationwide

Quantity

3044