FDA Recall Terminated

CyberKnife Robotic; Catalog/part number 032000 and 033000 Cosmetic cover package. Product Usage: The CyberKnife Robotic Radiosurgery System is indicated for treatment planning and image guided stereotactic Radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.

Recall: Z-0771-2016 · Initiated January 22, 2016

Recall

Recall Number
Z-0771-2016
Event Number
73193
Firm
Accuray Incorporated
FEI Number
1000123585
Product Code
IYE
Status
Terminated
Root Cause
Device Design
Initiated
January 22, 2016
Posted
February 9, 2016
Terminated
October 20, 2016
Address
1310 Chesapeake Ter, Sunnyvale, CA, 94089-1100

Description

CyberKnife Robotic; Catalog/part number 032000 and 033000 Cosmetic cover package. Product Usage: The CyberKnife Robotic Radiosurgery System is indicated for treatment planning and image guided stereotactic Radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.

Reason

The gun box mounting bracket may fail to support the weight of the gun box when in vertical (inverted) position. If this failure occurs the gun box may become loose and could come into contact with a patient.

Action

Accuray sent an Urgent Safety Advisory Notification letters dated January 22, 2016 to all affected customers by trackable mail method. The letters identified the affected product, problem and actions to be taken. Customers were instructed to complete and return the Acknowledgement form. For questions contact Customer Support at 1-877-668-8667 or visit the Accuray Technical Solution Center at http://www.accuray.com/services-support

Distribution

Worldwide Distribution. US Nationwide, Venezuela, United Kingdom, Ukraine, Turkey, Thailand, Taiwan, Switzerland, Spain, Saudi Arabia, Russia, Poland, Pakistan, Myanmar, Mexico, Korea, Japan, Italy, Ireland, India, Greece, Germany, France, Finland, Czech Republic, Colombia, China, Canada, and Belgium.

Quantity

226