FDA Recall Terminated

Trilogy F/M Acetabular Shell without holes. Several ODs: 40MM, 42MM, 44MM, 46MM, 48MM, 50MM, 52MM, 54MM, 56MM, 58MM, 60MM, 62MM, 64MM, 66MM, 68MM, 70MM. The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage.

Recall: Z-0769-2013 · Initiated September 14, 2012

Recall

Recall Number
Z-0769-2013
Event Number
63531
Firm
Zimmer, Inc.
FEI Number
1000220733
Product Code
LPH
Status
Terminated
Root Cause
Process control
Initiated
September 14, 2012
Posted
February 4, 2013
Terminated
June 20, 2014
Address
345 E Main St, Warsaw, IN, 46580-2746

Description

Trilogy F/M Acetabular Shell without holes. Several ODs: 40MM, 42MM, 44MM, 46MM, 48MM, 50MM, 52MM, 54MM, 56MM, 58MM, 60MM, 62MM, 64MM, 66MM, 68MM, 70MM. The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage.

Reason

Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy Acetabular System shells.

Action

Zimmer sent an "URGENT MEDICAL DEVICE NOTIFICATION" letter dated September 20, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.The firm expanded the scope of the recall in January 2013 to include devices that use a secondary mean of fixation. New recall letters will be submitted to distributors and surgeons.

Distribution

Worldwide Distribution-USA (nationwide) and the countries of Denmark, Germany, Greece, Hungary, Ireland, Italy, Netherlands, Spain, Sweden, UK, Poland, France, Australia, Chile, Canada, India, Japan, New Zealand, Peru, South Africa, Colombia, Ukraine, and UAE.

Quantity

275,952 units