FDA Recall Terminated

MicroScan LabPro Data Management System. Intended to manage both microbial identification (ID) and antimicrobial agent susceptibility testing (AST) data generated from MicroScan instruments or manually entered microbiology test results, for use by trained laboratory personnel.

Recall: Z-0768-2017 · Initiated October 25, 2016

Recall

Recall Number
Z-0768-2017
Event Number
75576
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
OUG
Status
Terminated
Root Cause
Device Design
Initiated
October 25, 2016
Posted
December 15, 2016
Terminated
May 1, 2017
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

MicroScan LabPro Data Management System. Intended to manage both microbial identification (ID) and antimicrobial agent susceptibility testing (AST) data generated from MicroScan instruments or manually entered microbiology test results, for use by trained laboratory personnel.

Reason

Beckman Coulter has received and confirmed reports of an intermittent and unexpected behavior when loading new panels into a WalkAway instrument using LabPro Data Management System version 4.42. The issue could cause workflow interruption with a potential of delaying reporting results due to the inability to begin processing new panels.

Action

An Urgent Medical Device Recall letter dated 10/25/16 was sent to all their customers to inform them that Beckman Coulter has received and confirmed reports of an intermittent and unexpected behavior when loading new panels into a WalkAway instrument using LabPro Data Management System version 4.42. Customers are informed that the issue could cause WalkAway workflow interruption with a potential of delaying reporting results due to the inability to begin processing new panels. The letter informs the customers of the actions to be taken and the resolution for the product on recall. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with any questions regarding the notice are instructed to contact their Customer Support representative: http://www.beckmancoulter.com or call 1-800-677-7226, option 1 in the United States.

Distribution

Worldwide Distribution -- United States and Mexico.

Quantity

131 units total (130 units in US)