FDA Recall Terminated

IVEA 500A Intended for Med-Surg patient care.

Recall: Z-0767-2017 · Initiated November 1, 2016

Recall

Recall Number
Z-0767-2017
Event Number
75571
Firm
Firefly Medical, Inc.
FEI Number
3011390406
Product Code
FOX
Status
Terminated
Root Cause
Process control
Initiated
November 1, 2016
Posted
December 15, 2016
Terminated
August 1, 2017
Address
320 E Vine Dr, Ste 312, Fort Collins, CO, 80524-2332

Description

IVEA 500A Intended for Med-Surg patient care.

Reason

The dowel pin could potentially slip out of its press fit, resulting in the product potentially slumping to one side which could cause the patient to stumble.

Action

Firm notified customers on **11/18/16** via certified "Urgent: Medical Device Recall (Field Correction) IVEA Model 500A" letter to correct product in the field. Customers were asked to thoroughly read the letter and to immediately follow the actions in the enclosed Product Correction Kit. The instructions include steps for cleaning the correction area and installing the permanent adhesive caps, as well as, completing and returning the product correction form to Firefly Medical. Customers may continue to use their units until the Product Correction actions are complete, as long as pins are not protruding. If the pins are protruding, then customers are to stop using the device immediately and complete the product correction instructions. Customers can call 970-472-5323 or e-mail [email protected] if they have any questions.

Distribution

Worldwide Distribution -- USA and Canada.

Quantity

239