FDA Recall Terminated

Philips Healthcare TraumaDiagnost; Codes: 72001, 720011, 720013, 720020; A radiological examination unit for emergency work and routine skeletal diagnosis.

Recall: Z-0766-2016 · Initiated July 22, 2015

Recall

Recall Number
Z-0766-2016
Event Number
71764
Firm
Philips Electronics North America Corporation
FEI Number
1218950
Product Code
KPR
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
July 22, 2015
Terminated
October 14, 2020
Address
3000 Minuteman Rd, Andover, MA, 01810-1032

Description

Philips Healthcare TraumaDiagnost; Codes: 72001, 720011, 720013, 720020; A radiological examination unit for emergency work and routine skeletal diagnosis.

Reason

If arm is not fastened securely it could fall down.

Action

Consignees were notified of the recall via letter sent certified return receipt on 7/22/2015. The letter informs customers that the Instruction for Use (IFU) requires the operator not to use the system and call for service, if a mechanical defect or malfunction is suspected. In this case, it is recommended not to position the Ceiling suspension with the Trauma arm directly above a patient as long as this Field Change Order is not yet installed. Philips plans to repair the affected systems. A Philips Service Engineer will contact customers as soon as the Field Action Kit is ready to be implemented. For questions, contact your local representative or use the Technical Support Line at 1-800-722-9377. When communicating with Philips regarding this issue, please reference Field Change Order 72000014.

Distribution

Distributed in the states of NY, WI, MA, AL, FL, GA, RI, CA, KS, and TX, and the countries of Canada, Australia, Austria, Belgium, France, Germany, Ireland, Italy, Netherlands, New Zealand, Portugal, South Africa, Spain, and Switzerland.

Quantity

399 distributed, 262 are subject to this event