P10 Tobii battery box (product number 510410) Product Usage - Battery pack for MyTobii P10 Device. MyTobii P10 is a portable eye-controlled communication device featuring the Tobii Eye Control system.
Recall
- Recall Number
- Z-0766-2013
- Event Number
- 64099
- Firm
- Tobii Assistive Technology, Inc.
- FEI Number
- 3007873415
- Product Code
- ILQ
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- January 14, 2013
- Posted
- February 1, 2013
- Terminated
- July 6, 2016
- Address
- 333 Elm St, Dedham, MA, 02026-4530
Description
P10 Tobii battery box (product number 510410) Product Usage - Battery pack for MyTobii P10 Device. MyTobii P10 is a portable eye-controlled communication device featuring the Tobii Eye Control system.
Tobii ATI customers P10 external battery pack, which contains sealed lead-acid batteries, malfunctioned while recharging and produced smoke with a chemical smell.
The recalling firm, Tobii Assistive Technology, Inc. sent a notification letter to all customers who received the recalled product via certified mail. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediatley discontinue the use of the P10 battery pack and provded with instructions to receive a complimentary replacement solution. The letter instructed customers to dispose of their P10 External Battery Pack in a manner stipulated with local laws. An acknowledgment form is included to be completed by customers. For questions contact Tobii ATI at 1-800-793-9227 or [email protected].
US Nationwide Distribution
483 units