FDA Recall Terminated

MicroSTAAR Injector w/ Foam Tip Plunger, Model MSI-PF

Recall: Z-0766-04 · Initiated February 23, 2004

Recall

Recall Number
Z-0766-04
Event Number
28570
Firm
Staar Surgical Co Inc
FEI Number
2023826
Product Code
KYB
Status
Terminated
Root Cause
Other
Initiated
February 23, 2004
Posted
July 20, 2004
Terminated
January 4, 2005
Address
1911 Walker Ave, Monrovia, CA, 91016-4846

Description

MicroSTAAR Injector w/ Foam Tip Plunger, Model MSI-PF

Reason

Injector mold design/manufacture and wear caused material flash to develop preventing assembly of the device prior to use.

Action

Firm will be sending thier representatives to customers to inspect inventory and arrange return of product.

Distribution

Nationwide.

Quantity

690 injectors