FDA Recall Open, Classified

Medline Fluid Delivery Set with Drip Chamber: 1) MTO,135,YELLOW,MICRO,SPK-MAIMONIDES, Item Number: 60130527; 2) MTO,ANGIO,DOUBLE,SPIKE,KIT,KENNEDY HLTH, Item Number: 65111341; 3) MTO,ANGIO,SINGLE,SPIKE,KIT, Item Number: 65232486; 4) MTO,ANGIO,SINGLE,SPIKE,KIT,GOOD,SAM, Item Number: 65071193; 5) MTO,NEURO,FLUSH,SPIKE,KIT,BLUE,FLORIDA, Item Number: 600604422; 6) MTO,NEURO,KIT,-,JERSEY,SHORE PG, Item Number: 60100045; 7) MTO,RADIOLOGY,KIT,HAMOT MED CEN, Item Number: 650801715; 8) MTS,135,CDS,IV,TUBE-JEFFERSON UNIV., Item Number: 60232481; 9) MTS,135,YELLOW,MICRO,SPIKE,LUTHERAN MED, Item Number: 60120583; 10) MTS,ANGIO,NEURO,KIT,NORTH SHORE UNIV, Item Number: 601403920; 11) MTS,CUST. FILTER-NEWYORK UNIVERSITY, Item Number: 62140602; 12) MTS,INTERV,NEURO,KIT-WESTCHESTER,MC, Item Number: 602301614; 13) MTS,NEU,NORTH SHORE HOSPITAL, Item Number: 6010771601; 14) MTS,NEURO,KIT, Item Number: 600602914; 15) MTS,NEURO,KIT,2,NEWENGLAND MED CTR, Item Number: 601400111; 16) MTS,NEURO,KIT,YLW&RED,FLORIDA HOSP, Item Number: 600604423; 17) MTS,NEURO,KIT-JOHN,F,KENNEDY MED CENTER, Item Number: 60110552; 18) MTS,NEURO,KIT-OVERLOOK HOSPITAL PG, Item Number: 60150035; 19) MTS,NEURO,SPIKE,KIT,CAPITAL HLTHSYS-FULD, Item Number: 60182723; 20) MTS,NEURO-RADIOLOGY,KIT-YALE-NEWHAVEN, Item Number: 602500214; 21) MTS,STROKE/EMBO,KIT,NORTH SHORE UNIV, Item Number: 601403921; 22) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601305210; 23) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601400110;

Recall: Z-0765-2025 · Initiated December 16, 2024

Recall

Recall Number
Z-0765-2025
Event Number
95988
Firm
MEDLINE INDUSTRIES, LP Northfield
FEI Number
1417592
Product Code
OEZ
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
December 16, 2024
Posted
January 15, 2025
Address
3 Lakes Dr, Northfield, IL, 60093-2753

Description

Medline Fluid Delivery Set with Drip Chamber: 1) MTO,135,YELLOW,MICRO,SPK-MAIMONIDES, Item Number: 60130527; 2) MTO,ANGIO,DOUBLE,SPIKE,KIT,KENNEDY HLTH, Item Number: 65111341; 3) MTO,ANGIO,SINGLE,SPIKE,KIT, Item Number: 65232486; 4) MTO,ANGIO,SINGLE,SPIKE,KIT,GOOD,SAM, Item Number: 65071193; 5) MTO,NEURO,FLUSH,SPIKE,KIT,BLUE,FLORIDA, Item Number: 600604422; 6) MTO,NEURO,KIT,-,JERSEY,SHORE PG, Item Number: 60100045; 7) MTO,RADIOLOGY,KIT,HAMOT MED CEN, Item Number: 650801715; 8) MTS,135,CDS,IV,TUBE-JEFFERSON UNIV., Item Number: 60232481; 9) MTS,135,YELLOW,MICRO,SPIKE,LUTHERAN MED, Item Number: 60120583; 10) MTS,ANGIO,NEURO,KIT,NORTH SHORE UNIV, Item Number: 601403920; 11) MTS,CUST. FILTER-NEWYORK UNIVERSITY, Item Number: 62140602; 12) MTS,INTERV,NEURO,KIT-WESTCHESTER,MC, Item Number: 602301614; 13) MTS,NEU,NORTH SHORE HOSPITAL, Item Number: 6010771601; 14) MTS,NEURO,KIT, Item Number: 600602914; 15) MTS,NEURO,KIT,2,NEWENGLAND MED CTR, Item Number: 601400111; 16) MTS,NEURO,KIT,YLW&RED,FLORIDA HOSP, Item Number: 600604423; 17) MTS,NEURO,KIT-JOHN,F,KENNEDY MED CENTER, Item Number: 60110552; 18) MTS,NEURO,KIT-OVERLOOK HOSPITAL PG, Item Number: 60150035; 19) MTS,NEURO,SPIKE,KIT,CAPITAL HLTHSYS-FULD, Item Number: 60182723; 20) MTS,NEURO-RADIOLOGY,KIT-YALE-NEWHAVEN, Item Number: 602500214; 21) MTS,STROKE/EMBO,KIT,NORTH SHORE UNIV, Item Number: 601403921; 22) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601305210; 23) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601400110;

Reason

Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.

Action

An IMMEDIATE ACTION REQUIRED MEDICAL DEVICE RECALL notification letter dated 12/16/24 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. The login for completing the response form is: Website link: https://recalls.medline.com Recall Reference #: R-24-270-FG Recall Code: 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected]. Update: A follow-up letter titled IMMEDIATE ACTION REQUIRED dated 2/11/25 was sent to customers. Medline would like to update our action to instruct you to please destroy all impacted convenience kits and Medline will provide replacement kit(s) with the correct Fluid Delivery Set. REQUIRED ACTION: 1. Immediately check your stock for the affected kit number(s) and lot number(s) listed on the attachment. Please destroy all affected kits. 2. Please complete and return the enclosed destruction form listing the quantity of affected kits destroyed. Even if you do not have any affected product, please complete and return the form. 3. Upon receipt of your completed destruction form, Medline will issue replacem

Distribution

US Nationwide Distribution

Quantity

3930 units