FDA Recall Open, Classified

Maquet PowerLED/HLED and PowerLED300 OR Light Systems, Model Numbers and Catalog Numbers EPS MB1 - ARDINT059002A, EPS MB1 - ARDINT059002C, EPS MB1 TR - ARDINT059000A, EPS MB1 TR - ARDINT059000C, EPS MB3 - ARDINT059003A, EPS MB3 - ARDINT059003C, EPS MB3 TR - ARDINT059001A, EPS MB3 TR - ARDINT059001C, OPT MBAT - ARD568210010A, OPT MBAT - ARD568210010C, PWD30 DUAL - ARD267800310C, PWD30 DUAL - ARD267800330A, PWD30 DUAL - ARD267800330C, PWD30 DUAL - ARD267800340A, PWD30 DUAL - ARD267800340C, PWD30+DF - ARD268300170A, PWD30+DF - ARD568424110A, PWD30+DF - ARD568424110C, PWD30+DF - ARD568424114A, PWD30+DF - ARD568424114C, PWD30+DF - ARD568424511A, PWD30+DF - ARD568424511C, PWD30+DF - ARD568424512A, PWD30+DF - ARD568424512C, PWD30+DF - ARD568424513A, PWD30+DF - ARD568424513C, PWD30+DF - ARD568484114A, PWD30+DF - ARD568484114C, PWD30+DF - ARD569424110A, PWD30+DF - ARD569424110C, PWD30+DF - ARD569424114A, PWD30+DF - ARD569424114C, PWD30+DF - ARD569424511A, PWD30+DF - ARD569424511C, PWD30+DF - ARD569424512A, PWD30+DF - ARD569424512C, PWD30+DF - ARD569424513A, PWD30+DF - ARD569424513C, PWD30+DF - ARD569484114A, PWD30+DF - ARD569484114C, PWD30+DF - ARDPWD208008A, PWD30+DF - ARDPWD208008C, PWD30+DF - ARDPWD209000A, PWD30+DF - ARDPWD209000C, PWD30+DF - ARDPWD209009A, PWD30+DF - ARDPWD209009C, PWD30+DF - ARDPWD239013A, PWD30+DF - ARDPWD239013C, PWD30+DF - ARDPWD239014A, PWD30+DF - ARDPWD239014C, PWD30+DF - ARDPWD239015A, PWD30+DF - ARDPWD239015C, PWD30+DF - ARDPWD239031A, PWD30+DF - ARDPWD239031C, PWD30+DF - ARDPWD239032A, PWD30+DF - ARDPWD239032C, PWD30+DF - ARDPWD239033A, PWD30+DF - ARDPWD239033C, PWD30+SF - ARD568414110A, PWD30+SF - ARD568414110C, PWD30+SF - ARD568414114A, PWD30+SF - ARD568414114C, PWD30+SF - ARD568414511A, PWD30+SF - ARD568414511C, PWD30+SF - ARD568414512A, PWD30+SF - ARD568414512C, PWD30+SF - ARD568414513A, PWD30+SF - ARD568414513C, PWD30+SF - ARD568474114A, PWD30+SF - ARD568474114C, PWD30+SF - ARD569414110A, PWD30+SF - ARD569414110C, PWD30+SF - ARD569414114A, PWD30+SF - ARD569414114C, PWD30+SF - ARD569414511A, PWD30+SF - ARD569414511C, PWD30+SF - ARD569414512A, PWD30+SF - ARD569414512C, PWD30+SF - ARD569414513A, PWD30+SF - ARD569414513C, PWD30+SF - ARD569474114A, PWD30+SF - ARD569474114C, PWD30+SF - ARDPWD208009A, PWD30+SF - ARDPWD208009C, PWD30+SF - ARDPWD209001A, PWD30+SF - ARDPWD209001C, PWD30+SF - ARDPWD209002A, PWD30+SF - ARDPWD209002C, PWD30+SF - ARDPWD209010A, PWD30+SF - ARDPWD209010C, PWD30+SF - ARDPWD209011A, PWD30+SF - ARDPWD209011C, PWD30+SF - ARDPWD239016A, PWD30+SF - ARDPWD239016C, PWD30+SF - ARDPWD239017A, PWD30+SF - ARDPWD239017C, PWD30+SF - ARDPWD239018A, PWD30+SF - ARDPWD239018C, PWD30+SF - ARDPWD239034A, PWD30+SF - ARDPWD239034C, PWD30+SF - ARDPWD239035A, PWD30+SF - ARDPWD239035C, PWD30+SF - ARDPWD239036A, PWD30+SF - ARDPWD239036C, PWD30+SHIP - ARD268400080A, PWD30+SHIP - ARD268400080C, PWD33 S - ARD267800200A, PWD33 S - ARD267800200C, PWD33+DF - ARD568424710A, PWD33+DF - ARD568424710C, PWD33+DF - ARD568424714A, PWD33+DF - ARD568424714C, PWD33+DF - ARD568424751A, PWD33+DF - ARD568424751C, PWD33+DF - ARD568424752A, PWD33+DF - ARD568424752C, PWD33+DF - ARD568424753A, PWD33+DF - ARD568424753C, PWD33+DF - ARD568424755A, PWD33+DF - ARD568424755C, PWD33+DF - ARD568424758A, PWD33+DF - ARD568424758C, PWD33+DF - ARD569424710A, PWD33+DF - ARD569424710C, PWD33+DF - ARD569424714A, PWD33+DF - ARD569424714C, PWD33+DF - ARD569424751A, PWD33+DF - ARD569424751C, PWD33+DF - ARD569424752A, PWD33+DF - ARD569424752C, PWD33+DF - ARD569424753A, PWD33+DF - ARD569424753C, PWD33+DF - ARD569424755A, PWD33+DF - ARD569424755C, PWD33+DF - ARD569424758A, PWD33+DF - ARD569424758C, PWD33+SF - ARD568414710A, PWD33+SF - ARD568414710C, PWD33+SF - ARD568414714A, PWD33+SF - ARD568414714C, PWD33+SF - ARD568414751A, PWD33+SF - ARD568414751C, PWD33+SF - ARD568414752A, PWD33+SF - ARD568414752C, PWD33+SF - ARD568414753A, PWD33+SF - ARD568414753C, PWD33+SF - ARD568414755A, PWD33+SF - ARD568414755C, PWD33+SF - ARD568414758A, PWD33+SF - ARD5684147

Recall: Z-0763-2024 · Initiated November 8, 2023

Recall

Recall Number
Z-0763-2024
Event Number
93567
Firm
Getinge Usa Sales Inc
FEI Number
3013876692
Product Code
FSY
Status
Open, Classified
Root Cause
Equipment maintenance
Initiated
November 8, 2023
Posted
January 19, 2024
Address
1 Geoffrey Way, Wayne, NJ, 07470-2035

Description

Maquet PowerLED/HLED and PowerLED300 OR Light Systems, Model Numbers and Catalog Numbers EPS MB1 - ARDINT059002A, EPS MB1 - ARDINT059002C, EPS MB1 TR - ARDINT059000A, EPS MB1 TR - ARDINT059000C, EPS MB3 - ARDINT059003A, EPS MB3 - ARDINT059003C, EPS MB3 TR - ARDINT059001A, EPS MB3 TR - ARDINT059001C, OPT MBAT - ARD568210010A, OPT MBAT - ARD568210010C, PWD30 DUAL - ARD267800310C, PWD30 DUAL - ARD267800330A, PWD30 DUAL - ARD267800330C, PWD30 DUAL - ARD267800340A, PWD30 DUAL - ARD267800340C, PWD30+DF - ARD268300170A, PWD30+DF - ARD568424110A, PWD30+DF - ARD568424110C, PWD30+DF - ARD568424114A, PWD30+DF - ARD568424114C, PWD30+DF - ARD568424511A, PWD30+DF - ARD568424511C, PWD30+DF - ARD568424512A, PWD30+DF - ARD568424512C, PWD30+DF - ARD568424513A, PWD30+DF - ARD568424513C, PWD30+DF - ARD568484114A, PWD30+DF - ARD568484114C, PWD30+DF - ARD569424110A, PWD30+DF - ARD569424110C, PWD30+DF - ARD569424114A, PWD30+DF - ARD569424114C, PWD30+DF - ARD569424511A, PWD30+DF - ARD569424511C, PWD30+DF - ARD569424512A, PWD30+DF - ARD569424512C, PWD30+DF - ARD569424513A, PWD30+DF - ARD569424513C, PWD30+DF - ARD569484114A, PWD30+DF - ARD569484114C, PWD30+DF - ARDPWD208008A, PWD30+DF - ARDPWD208008C, PWD30+DF - ARDPWD209000A, PWD30+DF - ARDPWD209000C, PWD30+DF - ARDPWD209009A, PWD30+DF - ARDPWD209009C, PWD30+DF - ARDPWD239013A, PWD30+DF - ARDPWD239013C, PWD30+DF - ARDPWD239014A, PWD30+DF - ARDPWD239014C, PWD30+DF - ARDPWD239015A, PWD30+DF - ARDPWD239015C, PWD30+DF - ARDPWD239031A, PWD30+DF - ARDPWD239031C, PWD30+DF - ARDPWD239032A, PWD30+DF - ARDPWD239032C, PWD30+DF - ARDPWD239033A, PWD30+DF - ARDPWD239033C, PWD30+SF - ARD568414110A, PWD30+SF - ARD568414110C, PWD30+SF - ARD568414114A, PWD30+SF - ARD568414114C, PWD30+SF - ARD568414511A, PWD30+SF - ARD568414511C, PWD30+SF - ARD568414512A, PWD30+SF - ARD568414512C, PWD30+SF - ARD568414513A, PWD30+SF - ARD568414513C, PWD30+SF - ARD568474114A, PWD30+SF - ARD568474114C, PWD30+SF - ARD569414110A, PWD30+SF - ARD569414110C, PWD30+SF - ARD569414114A, PWD30+SF - ARD569414114C, PWD30+SF - ARD569414511A, PWD30+SF - ARD569414511C, PWD30+SF - ARD569414512A, PWD30+SF - ARD569414512C, PWD30+SF - ARD569414513A, PWD30+SF - ARD569414513C, PWD30+SF - ARD569474114A, PWD30+SF - ARD569474114C, PWD30+SF - ARDPWD208009A, PWD30+SF - ARDPWD208009C, PWD30+SF - ARDPWD209001A, PWD30+SF - ARDPWD209001C, PWD30+SF - ARDPWD209002A, PWD30+SF - ARDPWD209002C, PWD30+SF - ARDPWD209010A, PWD30+SF - ARDPWD209010C, PWD30+SF - ARDPWD209011A, PWD30+SF - ARDPWD209011C, PWD30+SF - ARDPWD239016A, PWD30+SF - ARDPWD239016C, PWD30+SF - ARDPWD239017A, PWD30+SF - ARDPWD239017C, PWD30+SF - ARDPWD239018A, PWD30+SF - ARDPWD239018C, PWD30+SF - ARDPWD239034A, PWD30+SF - ARDPWD239034C, PWD30+SF - ARDPWD239035A, PWD30+SF - ARDPWD239035C, PWD30+SF - ARDPWD239036A, PWD30+SF - ARDPWD239036C, PWD30+SHIP - ARD268400080A, PWD30+SHIP - ARD268400080C, PWD33 S - ARD267800200A, PWD33 S - ARD267800200C, PWD33+DF - ARD568424710A, PWD33+DF - ARD568424710C, PWD33+DF - ARD568424714A, PWD33+DF - ARD568424714C, PWD33+DF - ARD568424751A, PWD33+DF - ARD568424751C, PWD33+DF - ARD568424752A, PWD33+DF - ARD568424752C, PWD33+DF - ARD568424753A, PWD33+DF - ARD568424753C, PWD33+DF - ARD568424755A, PWD33+DF - ARD568424755C, PWD33+DF - ARD568424758A, PWD33+DF - ARD568424758C, PWD33+DF - ARD569424710A, PWD33+DF - ARD569424710C, PWD33+DF - ARD569424714A, PWD33+DF - ARD569424714C, PWD33+DF - ARD569424751A, PWD33+DF - ARD569424751C, PWD33+DF - ARD569424752A, PWD33+DF - ARD569424752C, PWD33+DF - ARD569424753A, PWD33+DF - ARD569424753C, PWD33+DF - ARD569424755A, PWD33+DF - ARD569424755C, PWD33+DF - ARD569424758A, PWD33+DF - ARD569424758C, PWD33+SF - ARD568414710A, PWD33+SF - ARD568414710C, PWD33+SF - ARD568414714A, PWD33+SF - ARD568414714C, PWD33+SF - ARD568414751A, PWD33+SF - ARD568414751C, PWD33+SF - ARD568414752A, PWD33+SF - ARD568414752C, PWD33+SF - ARD568414753A, PWD33+SF - ARD568414753C, PWD33+SF - ARD568414755A, PWD33+SF - ARD568414755C, PWD33+SF - ARD568414758A, PWD33+SF - ARD5684147

Reason

A potential for a light system to fall in the operating room.

Action

An URGENT MEDICAL DEVICE CORRECTION letter dated 11/8/23 was sent to customers. Actions to be taken by the customer: Our records indicate that you have been identified as having purchased at least one model of one of these product ranges. Please identify your affected units using the referenced devices on page 1 and follow the instructions below. Please distribute this notice and forward the link below to access manuals to anyone within your organization who may need to use the manuals, and especially to your service or service provider who performs your maintenance. Light system fixing and other replacements: Regarding: - the fixing of the light systems: suspensions fixing screws, adapter fixing screws, bushing fixing screws - the brake screws - the safety segments - the batteries, a dedicated instruction to illustrate and facilitate understanding of what to replace and what to order for replacement has been issued. The instruction is available on a dedicated space on the web portal https://www.getinge.com/int/campaigns/maintenance-and-service-on-or-light-system/, and provided upon request by emailing [email protected] Should you have questions or require additional information, please do not hesitate to contact Maquet SAS/Getinge or your local representative. Please complete and sign the attached MEDICAL DEVICE CORRECTION - RESPONSE FORM (page 7) to acknowledge that you have received this notification. Return the completed form to Maquet/Getinge by e-mailing a scanned copy to [email protected] or by faxing the form to (866) 350-0897. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. Actions to be taken by Maquet SAS / Getinge: As the majority of customers may not have regular contact with Getinge services or authorized Getinge representatives and therefore may not be aware of current maintenance recommendations, Maquet SAS/Getinge is providin

Distribution

US Nationwide. Global Distribution.

Quantity

236793 units in total