FDA Recall Terminated

Heating Profile PRISMAFLO II S blood return warmer for Prismaflex Product Code: PF2-WP31 Product Usage: The PRISMAFLO IIS warmer is used for warming return blood flow.

Recall: Z-0760-2019 · Initiated September 6, 2018

Recall

Recall Number
Z-0760-2019
Event Number
81755
Firm
Stihler Electronic Gmbh Julius-Holder-Str. 36 Stuttgart Germany
FEI Number
3001733054
Product Code
KOC
Status
Terminated
Root Cause
Device Design
Initiated
September 6, 2018
Terminated
May 13, 2020

Description

Heating Profile PRISMAFLO II S blood return warmer for Prismaflex Product Code: PF2-WP31 Product Usage: The PRISMAFLO IIS warmer is used for warming return blood flow.

Reason

PRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn

Action

STIHLER ELECTRONIC GMBH notified distributors by Registered mail/email and to Head of Biomedical on 09/06/18 identifying the problem, health risk and action to be taken: identify the Heating Profiles according to the serial numbers specified. These profiles must be replaced based on your accurate reply. Complete Inventory must be accounted for. For your reply use attached "Return Form" to Stihler Electronic GmbH. Send the filled return form via e-mail to nicola. vosseler@sti hlerelectronic. de. 3. You will receive the replacement Heating Profiles from our distributor Futuremed America Inc., 15700 Devonshire Street, Granada Hills, 91344 California, US.

Distribution

Worldwide Distribution - US Nationwide in the states of CA and Foreign of: Switzerland

Quantity

1 unit