FDA Recall Open, Classified

Maquet Lucea - Lucea10/40, Lucea50/100 OR Light Systems, Model Numbers and Catalog Numbers LCA 10 - ARD2LCA00040C, LCA 10 - ARDLCA109000C, LCA 10 - ARDLCA109002C, LCA 10 - ARDLCA109006C, LCA 10 - ARDLCA209008A, LCA 10 - ARDLCA209008C, LCA 10 - ARDLCA209009A, LCA 10 - ARDLCA209009C, LCA 10 - ARDLCA209010A, LCA 10 - ARDLCA209010C, LCA 10 - ARDLCA409000C, LCA 10 - ARDLCA409001C, LCA 10 - ARDLCA409002A, LCA 10 - ARDLCA409002C, LCA 10 - ARDLCA409003C, LCA 10 - ARDLCA409004A, LCA 10 - ARDLCA409004C, LCA 10 - ARDLCA409005C, LCA 10 DES - ARDLCA409001A, LCA 10 MOBILE - ARDLCA309008A, LCA 10 RAIL - ARDLCA409005A, LCA 10 WALL - ARDLCA109006A, LCA 40 - ARDLCA109001C, LCA 40 - ARDLCA208000A, LCA 40 - ARDLCA208000C, LCA 40 - ARDLCA208001A, LCA 40 - ARDLCA208001C, LCA 40 - ARDLCA209000A, LCA 40 - ARDLCA209000C, LCA 40 - ARDLCA209001A, LCA 40 - ARDLCA209001C, LCA 40 - ARDLCA209002A, LCA 40 - ARDLCA209002C, LCA 40 - ARDLCA209003A, LCA 40 - ARDLCA209003C, LCA 40 - ARDLCA209014A, LCA 40 - ARDLCA209014C, LCA 40 - ARDLCA209017A, LCA 40 - ARDLCA209017C, LCA 40 - ARDLCA209018A, LCA 40 - ARDLCA209018C, LCA 40 - ARDLCA209019A, LCA 40 - ARDLCA209019C, LCA 40 - ARDLCA209020A, LCA 40 - ARDLCA209020C, LCA 40 - ARDLCA209021A, LCA 40 - ARDLCA209021C, LCA 40 MOBILE - ARDLCA309009A, LCA 40 WALL - ARDLCA109001A, LCA 40GS - ARDLCA109009A, LCA 40GS - ARDLCA109009C, LCA 100 - ARDLCA109005A, LCA 100 - ARDLCA109005C, LCA 100 - ARDLCA219000A, LCA 100 MOBILE - ARDLCA309006A, LCA 100 MOBILE B - ARDLCA309007A, LCA 100 SAT - ARDSAT209004A, LCA 100 SAT - ARDSAT209005A, LCA 100 V - ARDLCA219000C, LCA 100DF - ARDLCA109008A, LCA 100DF - ARDLCA109008C, LCA 100DF - ARDLCA219003A, LCA 100DF - ARDSAT209008A, LCA 100DF - ARDSAT209009A, LCA 100DF V - ARDLCA219003C, LCA 50 - ARDLCA109004A, LCA 50 - ARDLCA109004C, LCA 50 - ARDLCA209012A, LCA 50 - ARDLCA209012C, LCA 50 - ARDSAT209002A, LCA 50 MOBILE - ARDLCA309004A, LCA 50 MOBILE B - ARDLCA309005A, LCA 50 SAT - ARDSAT209003A, LCA 50 WALL 11 - ARD2LCA00140A, LCA 50DF - ARDLCA109007A, LCA 50DF - ARDLCA109007C, LCA 50DF - ARDLCA209015A, LCA 50DF - ARDLCA209015C, LCA 50DF - ARDSAT209006A, LCA 50DF - ARDSAT209007A, LCA DUO 100 - ARDLCA219001A, LCA DUO 100 V - ARDLCA219001C, LCA DUO 100+50 - ARDLCA219002A, LCA DUO 100+50 V - ARDLCA219002C, LCA DUO 100+50DF - ARDLCA219005A, LCA DUO 100+50DF V - ARDLCA219005C, LCA DUO 100DF - ARDLCA219004A, LCA DUO 100DF V - ARDLCA219004C, LCA DUO 50 - ARDLCA209013A, LCA DUO 50 - ARDLCA209013C, LCA DUO 50DF - ARDLCA209016A, LCA DUO 50DF - ARDLCA209016C, LCA100 MOBILE UPGRADE - ARDLCA506006A, LCA100 UPGRADE - ARDLCA506004A, LCA100 UPGRADE 24V - ARDLCA506005A, LCA100 WALL UPGRADE - ARDLCA506007A, LCA50 UPGRADE - ARDLCA506000A, LCA50 UPGRADE 24V - ARDLCA506001A, LCA50 UPGRADE WALL MOUNTED - ARDLCA506003A, SAT LCA 100 - ARDSAT209004C, SAT LCA 100 - ARDSAT209005C, SAT LCA 100DF - ARDSAT209008C, SAT LCA 100DF - ARDSAT209009C, SAT LCA 50 - ARDSAT209002C, SAT LCA 50 - ARDSAT209003C, SAT LCA 50DF - ARDSAT209006C, SAT LCA 50DF - ARDSAT209007C, LCA 10 MOB - ARDLCA309000A, LCA 10 MOB - ARDLCA309000C, LCA 10 MOB - ARDLCA309002A, LCA 10 MOB - ARDLCA309002C, LCA 10 MOB - ARDLCA309008C, LCA 100 MOBILE - ARDLCA309006C, LCA 100 MOBILE B - ARDLCA309007C, LCA 40 MOB - ARDLCA309001A, LCA 40 MOB - ARDLCA309001C, LCA 40 MOB - ARDLCA309003A, LCA 40 MOB - ARDLCA309003C, LCA 40 MOB - ARDLCA309009C, LCA 40GS MOBILE - ARDLCA309010A, LCA 40GS MOBILE - ARDLCA309010C, LCA 50 MOBILE - ARDLCA309004C, LCA 50 MOBILE B - ARDLCA309005C,

Recall: Z-0759-2024 · Initiated November 8, 2023

Recall

Recall Number
Z-0759-2024
Event Number
93567
Firm
Getinge Usa Sales Inc
FEI Number
3013876692
Product Code
FTD
Status
Open, Classified
Root Cause
Equipment maintenance
Initiated
November 8, 2023
Posted
January 19, 2024
Address
1 Geoffrey Way, Wayne, NJ, 07470-2035

Description

Maquet Lucea - Lucea10/40, Lucea50/100 OR Light Systems, Model Numbers and Catalog Numbers LCA 10 - ARD2LCA00040C, LCA 10 - ARDLCA109000C, LCA 10 - ARDLCA109002C, LCA 10 - ARDLCA109006C, LCA 10 - ARDLCA209008A, LCA 10 - ARDLCA209008C, LCA 10 - ARDLCA209009A, LCA 10 - ARDLCA209009C, LCA 10 - ARDLCA209010A, LCA 10 - ARDLCA209010C, LCA 10 - ARDLCA409000C, LCA 10 - ARDLCA409001C, LCA 10 - ARDLCA409002A, LCA 10 - ARDLCA409002C, LCA 10 - ARDLCA409003C, LCA 10 - ARDLCA409004A, LCA 10 - ARDLCA409004C, LCA 10 - ARDLCA409005C, LCA 10 DES - ARDLCA409001A, LCA 10 MOBILE - ARDLCA309008A, LCA 10 RAIL - ARDLCA409005A, LCA 10 WALL - ARDLCA109006A, LCA 40 - ARDLCA109001C, LCA 40 - ARDLCA208000A, LCA 40 - ARDLCA208000C, LCA 40 - ARDLCA208001A, LCA 40 - ARDLCA208001C, LCA 40 - ARDLCA209000A, LCA 40 - ARDLCA209000C, LCA 40 - ARDLCA209001A, LCA 40 - ARDLCA209001C, LCA 40 - ARDLCA209002A, LCA 40 - ARDLCA209002C, LCA 40 - ARDLCA209003A, LCA 40 - ARDLCA209003C, LCA 40 - ARDLCA209014A, LCA 40 - ARDLCA209014C, LCA 40 - ARDLCA209017A, LCA 40 - ARDLCA209017C, LCA 40 - ARDLCA209018A, LCA 40 - ARDLCA209018C, LCA 40 - ARDLCA209019A, LCA 40 - ARDLCA209019C, LCA 40 - ARDLCA209020A, LCA 40 - ARDLCA209020C, LCA 40 - ARDLCA209021A, LCA 40 - ARDLCA209021C, LCA 40 MOBILE - ARDLCA309009A, LCA 40 WALL - ARDLCA109001A, LCA 40GS - ARDLCA109009A, LCA 40GS - ARDLCA109009C, LCA 100 - ARDLCA109005A, LCA 100 - ARDLCA109005C, LCA 100 - ARDLCA219000A, LCA 100 MOBILE - ARDLCA309006A, LCA 100 MOBILE B - ARDLCA309007A, LCA 100 SAT - ARDSAT209004A, LCA 100 SAT - ARDSAT209005A, LCA 100 V - ARDLCA219000C, LCA 100DF - ARDLCA109008A, LCA 100DF - ARDLCA109008C, LCA 100DF - ARDLCA219003A, LCA 100DF - ARDSAT209008A, LCA 100DF - ARDSAT209009A, LCA 100DF V - ARDLCA219003C, LCA 50 - ARDLCA109004A, LCA 50 - ARDLCA109004C, LCA 50 - ARDLCA209012A, LCA 50 - ARDLCA209012C, LCA 50 - ARDSAT209002A, LCA 50 MOBILE - ARDLCA309004A, LCA 50 MOBILE B - ARDLCA309005A, LCA 50 SAT - ARDSAT209003A, LCA 50 WALL 11 - ARD2LCA00140A, LCA 50DF - ARDLCA109007A, LCA 50DF - ARDLCA109007C, LCA 50DF - ARDLCA209015A, LCA 50DF - ARDLCA209015C, LCA 50DF - ARDSAT209006A, LCA 50DF - ARDSAT209007A, LCA DUO 100 - ARDLCA219001A, LCA DUO 100 V - ARDLCA219001C, LCA DUO 100+50 - ARDLCA219002A, LCA DUO 100+50 V - ARDLCA219002C, LCA DUO 100+50DF - ARDLCA219005A, LCA DUO 100+50DF V - ARDLCA219005C, LCA DUO 100DF - ARDLCA219004A, LCA DUO 100DF V - ARDLCA219004C, LCA DUO 50 - ARDLCA209013A, LCA DUO 50 - ARDLCA209013C, LCA DUO 50DF - ARDLCA209016A, LCA DUO 50DF - ARDLCA209016C, LCA100 MOBILE UPGRADE - ARDLCA506006A, LCA100 UPGRADE - ARDLCA506004A, LCA100 UPGRADE 24V - ARDLCA506005A, LCA100 WALL UPGRADE - ARDLCA506007A, LCA50 UPGRADE - ARDLCA506000A, LCA50 UPGRADE 24V - ARDLCA506001A, LCA50 UPGRADE WALL MOUNTED - ARDLCA506003A, SAT LCA 100 - ARDSAT209004C, SAT LCA 100 - ARDSAT209005C, SAT LCA 100DF - ARDSAT209008C, SAT LCA 100DF - ARDSAT209009C, SAT LCA 50 - ARDSAT209002C, SAT LCA 50 - ARDSAT209003C, SAT LCA 50DF - ARDSAT209006C, SAT LCA 50DF - ARDSAT209007C, LCA 10 MOB - ARDLCA309000A, LCA 10 MOB - ARDLCA309000C, LCA 10 MOB - ARDLCA309002A, LCA 10 MOB - ARDLCA309002C, LCA 10 MOB - ARDLCA309008C, LCA 100 MOBILE - ARDLCA309006C, LCA 100 MOBILE B - ARDLCA309007C, LCA 40 MOB - ARDLCA309001A, LCA 40 MOB - ARDLCA309001C, LCA 40 MOB - ARDLCA309003A, LCA 40 MOB - ARDLCA309003C, LCA 40 MOB - ARDLCA309009C, LCA 40GS MOBILE - ARDLCA309010A, LCA 40GS MOBILE - ARDLCA309010C, LCA 50 MOBILE - ARDLCA309004C, LCA 50 MOBILE B - ARDLCA309005C,

Reason

A potential for a light system to fall in the operating room.

Action

An URGENT MEDICAL DEVICE CORRECTION letter dated 11/8/23 was sent to customers. Actions to be taken by the customer: Our records indicate that you have been identified as having purchased at least one model of one of these product ranges. Please identify your affected units using the referenced devices on page 1 and follow the instructions below. Please distribute this notice and forward the link below to access manuals to anyone within your organization who may need to use the manuals, and especially to your service or service provider who performs your maintenance. Light system fixing and other replacements: Regarding: - the fixing of the light systems: suspensions fixing screws, adapter fixing screws, bushing fixing screws - the brake screws - the safety segments - the batteries, a dedicated instruction to illustrate and facilitate understanding of what to replace and what to order for replacement has been issued. The instruction is available on a dedicated space on the web portal https://www.getinge.com/int/campaigns/maintenance-and-service-on-or-light-system/, and provided upon request by emailing [email protected] Should you have questions or require additional information, please do not hesitate to contact Maquet SAS/Getinge or your local representative. Please complete and sign the attached MEDICAL DEVICE CORRECTION - RESPONSE FORM (page 7) to acknowledge that you have received this notification. Return the completed form to Maquet/Getinge by e-mailing a scanned copy to [email protected] or by faxing the form to (866) 350-0897. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. Actions to be taken by Maquet SAS / Getinge: As the majority of customers may not have regular contact with Getinge services or authorized Getinge representatives and therefore may not be aware of current maintenance recommendations, Maquet SAS/Getinge is providin

Distribution

US Nationwide. Global Distribution.

Quantity

236793 units in total