FDA Recall Open, Classified

Maquet Volista, Volista Access, Volista Access II, Volista Standop OR Light System, Model Numbers and Catalog Numbers ROL B FOR VLT400SF STP - ARDROL309001A, ROL B FOR VLT400SF STP - ARDROL309001C, ROL B FOR VLT600SF STP - ARDROL309003A, ROL B FOR VLT600SF STP - ARDROL309003C, ROL FOR VLT400SF STP - ARDROL309000A, ROL FOR VLT400SF STP - ARDROL309000C, ROL FOR VLT600SF STP - ARDROL309002A, ROL FOR VLT600SF STP - ARDROL309002C, VCS40 ROL - ARDROL309024A, VCS40 ROL - ARDROL309024C, VCS40 ROL - ARDROL309028A, VCS40 ROL - ARDROL309028C, VCS60 ROL - ARDROL309025A, VCS60 ROL - ARDROL309025C, VCS60 ROL - ARDROL309029A, VCS60 ROL - ARDROL309029C, VST40 - ARDROL309016A, VST40 - ARDROL309016C, VST40 - ARDROL309017A, VST40 - ARDROL309017C, VST40 - ARDROL309020A, VST40 - ARDROL309020C, VST40 - ARDROL309022A, VST40 - ARDROL309022C, VST60 - ARDROL309018A, VST60 - ARDROL309018C, VST60 - ARDROL309019A, VST60 - ARDROL309019C, VST60 - ARDROL309021A, VST60 - ARDROL309021C, VST60 - ARDROL309023A, VST60 - ARDROL309023C, VCS40DF - ARDVCS209001A, VCS40DF - ARDVCS209001C, VCS40DF - ARDVCS209010A, VCS40DF - ARDVCS209012A, VCS40DF - ARDVCS209021A, VCS40SF - ARDVCS209000A, VCS40SF - ARDVCS209000C, VCS40SF - ARDVCS209011A, VCS40SF 08 - ARDVCS209046A, VCS44DF - ARDVCS209005A, VCS44DF - ARDVCS209005C, VCS44DF - ARDVCS209016A, VCS44SF - ARDVCS209004A, VCS44SF - ARDVCS209004C, VCS44SF - ARDVCS209015A, VCS44SFDF - ARD2VCS00040A, VCS60DF - ARDVCS209003A, VCS60DF - ARDVCS209003C, VCS60DF - ARDVCS209014A, VCS60DF - ARDVCS209022A, VCS60DF 11 - ARDVCS209047A, VCS60SF - ARDVCS209002A, VCS60SF - ARDVCS209002C, VCS60SF - ARDVCS209013A, VCS64DF - ARDVCS209007A, VCS64DF - ARDVCS209007C, VCS64DF - ARDVCS209018A, VCS64SF - ARDVCS209006A, VCS64SF - ARDVCS209006C, VCS64SF - ARDVCS209017A, VCS64SFDF - ARD2VCS00010A, VCS64SFDF 1011 - ARDVCS209048A, VCS66DF - ARDVCS209009A, VCS66DF - ARDVCS209009C, VCS66DF - ARDVCS209020A, VCS66SF - ARDVCS209008A, VCS66SF - ARDVCS209008C, VCS66SF - ARDVCS209019A, VCSII40DF SB08 - ARDVCS209023A, VCSII40DF SB11 - ARDVCS209024A, VCSII40SF SB11 - ARDVCS209025A, VCSII44DF SB0810 - ARDVCS209026A, VCSII44SF SB1011 - ARDVCS209027A, VCSII60DF SB08 - ARDVCS209028A, VCSII60SF SB11 - ARDVCS209029A, VCSII64DF SB0810 - ARDVCS209030A, VCSII64SF SB1011 - ARDVCS209031A, VCSII66DF SB0810 - ARDVCS209032A, VCSII66SF SB1011 - ARDVCS209033A, VCSII40DF TK SB08 - ARDVCS209034A, VCSII40DF TK SB11 - ARDVCS209035A, VCSII40SF TK SB11 - ARDVCS209036A, VCSII44DF TK SB0810 - ARDVCS209037A, VCSII44SF TK SB1011 - ARDVCS209038A, VCSII60DF TK SB08 - ARDVCS209039A, VCSII60SF TK SB11 - ARDVCS209040A, VCSII64DF TK SB0810 - ARDVCS209041A, VCSII64SF TK SB1011 - ARDVCS209042A, VCSII66DF TK SB0810 - ARDVCS209043A, VCSII66SF TK SB1011 - ARDVCS209044A, VCS40DF SAT 12 - ARD2VCS00070A, VCSII40SF SB08 - ARDVCS209049A, VCSII64SFDF SB1011 - ARDVCS209050A, SAT VLT400D - ARD2SAT00080C, SAT VLT400D - ARD2SAT00100C, SAT VLT400DF - ARD2SAT00080A, SAT VLT400DF - ARD2SAT00100A, SAT VLT400S - ARD2SAT00030C, SAT VLT400S - ARD2SAT00040C, SAT VLT400SF - ARD2SAT00040A, SAT VLT40DF - ARD268800130A, SAT VLT40DF - ARD268800130C, SAT VLT40SF - ARD268800120A, SAT VLT40SF - ARD268800120C, SAT VLT60DF - ARD2SAT00170A, SAT VLT60DF - ARD2SAT00170C, SAT VLT60SF - ARD2SAT00180C, SAT VST40DF - ARDSAT209018A, SAT VST40DF - ARDSAT209019A, SAT VST40DF - ARDSAT209022A, SAT VST40DF - ARDSAT209023A, SAT VST40SF - ARDSAT209010A, SAT VST40SF - ARDSAT209011A, SAT VST40SF - ARDSAT209014A, SAT VST40SF - ARDSAT209015A, SAT VST60DF - ARDSAT209034A, SAT VST60DF - ARDSAT209035A, SAT VST60DF - ARDSAT209038A, SAT VST60DF - ARDSAT209039A, SAT VST60SF - ARDSAT209026A, SAT VST60SF - ARDSAT209027A, SAT VST60SF - ARDSAT209030A, SAT VST60SF - ARDSAT209031A, STANDOP 40DF SINGLE - ARDSTP229005A, STANDOP 40SF SINGLE - ARDSTP229000A, STANDOP 44DF DOUBLE - ARDSTP229007A, STANDOP 44SF DOUBLE - ARDSTP229002A, STANDOP 60DF SINGLE - ARDSTP229006A, STANDOP 60SF SINGLE - ARDSTP229001A, STANDOP 64DF DOUBLE - ARDSTP229008A, STANDOP 64SF DOUBLE - ARDS

Recall: Z-0758-2024 · Initiated November 8, 2023

Recall

Recall Number
Z-0758-2024
Event Number
93567
Firm
Getinge Usa Sales Inc
FEI Number
3013876692
Product Code
FTD
Status
Open, Classified
Root Cause
Equipment maintenance
Initiated
November 8, 2023
Posted
January 19, 2024
Address
1 Geoffrey Way, Wayne, NJ, 07470-2035

Description

Maquet Volista, Volista Access, Volista Access II, Volista Standop OR Light System, Model Numbers and Catalog Numbers ROL B FOR VLT400SF STP - ARDROL309001A, ROL B FOR VLT400SF STP - ARDROL309001C, ROL B FOR VLT600SF STP - ARDROL309003A, ROL B FOR VLT600SF STP - ARDROL309003C, ROL FOR VLT400SF STP - ARDROL309000A, ROL FOR VLT400SF STP - ARDROL309000C, ROL FOR VLT600SF STP - ARDROL309002A, ROL FOR VLT600SF STP - ARDROL309002C, VCS40 ROL - ARDROL309024A, VCS40 ROL - ARDROL309024C, VCS40 ROL - ARDROL309028A, VCS40 ROL - ARDROL309028C, VCS60 ROL - ARDROL309025A, VCS60 ROL - ARDROL309025C, VCS60 ROL - ARDROL309029A, VCS60 ROL - ARDROL309029C, VST40 - ARDROL309016A, VST40 - ARDROL309016C, VST40 - ARDROL309017A, VST40 - ARDROL309017C, VST40 - ARDROL309020A, VST40 - ARDROL309020C, VST40 - ARDROL309022A, VST40 - ARDROL309022C, VST60 - ARDROL309018A, VST60 - ARDROL309018C, VST60 - ARDROL309019A, VST60 - ARDROL309019C, VST60 - ARDROL309021A, VST60 - ARDROL309021C, VST60 - ARDROL309023A, VST60 - ARDROL309023C, VCS40DF - ARDVCS209001A, VCS40DF - ARDVCS209001C, VCS40DF - ARDVCS209010A, VCS40DF - ARDVCS209012A, VCS40DF - ARDVCS209021A, VCS40SF - ARDVCS209000A, VCS40SF - ARDVCS209000C, VCS40SF - ARDVCS209011A, VCS40SF 08 - ARDVCS209046A, VCS44DF - ARDVCS209005A, VCS44DF - ARDVCS209005C, VCS44DF - ARDVCS209016A, VCS44SF - ARDVCS209004A, VCS44SF - ARDVCS209004C, VCS44SF - ARDVCS209015A, VCS44SFDF - ARD2VCS00040A, VCS60DF - ARDVCS209003A, VCS60DF - ARDVCS209003C, VCS60DF - ARDVCS209014A, VCS60DF - ARDVCS209022A, VCS60DF 11 - ARDVCS209047A, VCS60SF - ARDVCS209002A, VCS60SF - ARDVCS209002C, VCS60SF - ARDVCS209013A, VCS64DF - ARDVCS209007A, VCS64DF - ARDVCS209007C, VCS64DF - ARDVCS209018A, VCS64SF - ARDVCS209006A, VCS64SF - ARDVCS209006C, VCS64SF - ARDVCS209017A, VCS64SFDF - ARD2VCS00010A, VCS64SFDF 1011 - ARDVCS209048A, VCS66DF - ARDVCS209009A, VCS66DF - ARDVCS209009C, VCS66DF - ARDVCS209020A, VCS66SF - ARDVCS209008A, VCS66SF - ARDVCS209008C, VCS66SF - ARDVCS209019A, VCSII40DF SB08 - ARDVCS209023A, VCSII40DF SB11 - ARDVCS209024A, VCSII40SF SB11 - ARDVCS209025A, VCSII44DF SB0810 - ARDVCS209026A, VCSII44SF SB1011 - ARDVCS209027A, VCSII60DF SB08 - ARDVCS209028A, VCSII60SF SB11 - ARDVCS209029A, VCSII64DF SB0810 - ARDVCS209030A, VCSII64SF SB1011 - ARDVCS209031A, VCSII66DF SB0810 - ARDVCS209032A, VCSII66SF SB1011 - ARDVCS209033A, VCSII40DF TK SB08 - ARDVCS209034A, VCSII40DF TK SB11 - ARDVCS209035A, VCSII40SF TK SB11 - ARDVCS209036A, VCSII44DF TK SB0810 - ARDVCS209037A, VCSII44SF TK SB1011 - ARDVCS209038A, VCSII60DF TK SB08 - ARDVCS209039A, VCSII60SF TK SB11 - ARDVCS209040A, VCSII64DF TK SB0810 - ARDVCS209041A, VCSII64SF TK SB1011 - ARDVCS209042A, VCSII66DF TK SB0810 - ARDVCS209043A, VCSII66SF TK SB1011 - ARDVCS209044A, VCS40DF SAT 12 - ARD2VCS00070A, VCSII40SF SB08 - ARDVCS209049A, VCSII64SFDF SB1011 - ARDVCS209050A, SAT VLT400D - ARD2SAT00080C, SAT VLT400D - ARD2SAT00100C, SAT VLT400DF - ARD2SAT00080A, SAT VLT400DF - ARD2SAT00100A, SAT VLT400S - ARD2SAT00030C, SAT VLT400S - ARD2SAT00040C, SAT VLT400SF - ARD2SAT00040A, SAT VLT40DF - ARD268800130A, SAT VLT40DF - ARD268800130C, SAT VLT40SF - ARD268800120A, SAT VLT40SF - ARD268800120C, SAT VLT60DF - ARD2SAT00170A, SAT VLT60DF - ARD2SAT00170C, SAT VLT60SF - ARD2SAT00180C, SAT VST40DF - ARDSAT209018A, SAT VST40DF - ARDSAT209019A, SAT VST40DF - ARDSAT209022A, SAT VST40DF - ARDSAT209023A, SAT VST40SF - ARDSAT209010A, SAT VST40SF - ARDSAT209011A, SAT VST40SF - ARDSAT209014A, SAT VST40SF - ARDSAT209015A, SAT VST60DF - ARDSAT209034A, SAT VST60DF - ARDSAT209035A, SAT VST60DF - ARDSAT209038A, SAT VST60DF - ARDSAT209039A, SAT VST60SF - ARDSAT209026A, SAT VST60SF - ARDSAT209027A, SAT VST60SF - ARDSAT209030A, SAT VST60SF - ARDSAT209031A, STANDOP 40DF SINGLE - ARDSTP229005A, STANDOP 40SF SINGLE - ARDSTP229000A, STANDOP 44DF DOUBLE - ARDSTP229007A, STANDOP 44SF DOUBLE - ARDSTP229002A, STANDOP 60DF SINGLE - ARDSTP229006A, STANDOP 60SF SINGLE - ARDSTP229001A, STANDOP 64DF DOUBLE - ARDSTP229008A, STANDOP 64SF DOUBLE - ARDS

Reason

A potential for a light system to fall in the operating room.

Action

An URGENT MEDICAL DEVICE CORRECTION letter dated 11/8/23 was sent to customers. Actions to be taken by the customer: Our records indicate that you have been identified as having purchased at least one model of one of these product ranges. Please identify your affected units using the referenced devices on page 1 and follow the instructions below. Please distribute this notice and forward the link below to access manuals to anyone within your organization who may need to use the manuals, and especially to your service or service provider who performs your maintenance. Light system fixing and other replacements: Regarding: - the fixing of the light systems: suspensions fixing screws, adapter fixing screws, bushing fixing screws - the brake screws - the safety segments - the batteries, a dedicated instruction to illustrate and facilitate understanding of what to replace and what to order for replacement has been issued. The instruction is available on a dedicated space on the web portal https://www.getinge.com/int/campaigns/maintenance-and-service-on-or-light-system/, and provided upon request by emailing [email protected] Should you have questions or require additional information, please do not hesitate to contact Maquet SAS/Getinge or your local representative. Please complete and sign the attached MEDICAL DEVICE CORRECTION - RESPONSE FORM (page 7) to acknowledge that you have received this notification. Return the completed form to Maquet/Getinge by e-mailing a scanned copy to [email protected] or by faxing the form to (866) 350-0897. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. Actions to be taken by Maquet SAS / Getinge: As the majority of customers may not have regular contact with Getinge services or authorized Getinge representatives and therefore may not be aware of current maintenance recommendations, Maquet SAS/Getinge is providin

Distribution

US Nationwide. Global Distribution.

Quantity

236793 units in total