Maquet PowerLEDII OR Light System, Model Names and Catalog Numbers PWDII77SF - ARDPWT629100A, PWDII555DF - ARDPWT629101A, PWDII555SF - ARDPWT629102A, PWDII755DF - ARDPWT629103A, PWDII755SF - ARDPWT629104A, PWDII775DF - ARDPWT629105A, PWDII775SF - ARDPWT629106A, PWDII50DF - ARDPWT629109A, PWDII50SF - ARDPWT629110A, PWDII50SF - ARDPWT629111A, PWDII55DF - ARDPWT629112A, PWDII55SF - ARDPWT629113A, PWDII70DF - ARDPWT629114A, PWDII70DF - ARDPWT629115A, PWDII70SF - ARDPWT629116A, PWDII70SF - ARDPWT629117A, PWDII75DF - ARDPWT629118A, PWDII75SF - ARDPWT629119A, PWDII50DF - ARDPWT629120A, PWDII50SF - ARDPWT629121A, PWDII50SF - ARDPWT629122A, PWDII55DF - ARDPWT629123A, PWDII55SF - ARDPWT629124A, PWDII70DF - ARDPWT629125A, PWDII70DF - ARDPWT629126A, PWDII70SF - ARDPWT629127A, PWDII70SF - ARDPWT629128A, PWDII75DF - ARDPWT629129A, PWDII75SF - ARDPWT629130A, PWDII50DF - ARDPWT629131A, PWDII50SF - ARDPWT629132A, PWDII50SF - ARDPWT629133A, PWDII55DF - ARDPWT629134A, PWDII55SF - ARDPWT629135A, PWDII70DF - ARDPWT629136A, PWDII70DF - ARDPWT629137A, PWDII70SF - ARDPWT629138A, PWDII70SF - ARDPWT629139A, PWDII75DF - ARDPWT629140A, PWDII75SF - ARDPWT629141A, PWDII50DF - ARDPWT629142A, PWDII50SF - ARDPWT629143A, PWDII50SF - ARDPWT629144A, PWDII55DF - ARDPWT629145A, PWDII55SF - ARDPWT629146A, PWDII70DF - ARDPWT629147A, PWDII70DF - ARDPWT629148A, PWDII70SF - ARDPWT629149A, PWDII70SF - ARDPWT629150A, PWDII75DF - ARDPWT629151A, PWDII75SF - ARDPWT629152A, PWDII50DF - ARDPWT629154A, PWDII50DF - ARDPWT629155A, PWDII77SF - ARDPWT629156A, PWDII50DF - ARDPWT639001A, PWDII50DF - ARDPWT639002A, PWDII50DF - ARDPWT639003A, PWDII50DF - ARDPWT639004A, PWDII50SF - ARDPWT639006A, PWDII50SF - ARDPWT639007A, PWDII50SF - ARDPWT639008A, PWDII50SF - ARDPWT639009A, PWDII70DF - ARDPWT639011A, PWDII70DF - ARDPWT639012A, PWDII70DF - ARDPWT639013A, PWDII70DF - ARDPWT639014A, PWDII70SF - ARDPWT639016A, PWDII70SF - ARDPWT639017A, PWDII70SF - ARDPWT639018A, PWDII70SF - ARDPWT639019A, PWDII70DF - ARDPWT639021A, PWDII50DF - ARDPWT639022A, PWDII50DF - ARDPWT639023A, PWDII50DF - ARDPWT639024A, PWDII50DF - ARDPWT639025A, PWDII50SF - ARDPWT639027A, PWDII50SF - ARDPWT639028A, PWDII50SF - ARDPWT639029A, PWDII50SF - ARDPWT639030A, PWDII70DF - ARDPWT639032A, PWDII70DF - ARDPWT639033A, PWDII70DF - ARDPWT639034A, PWDII70DF - ARDPWT639035A, PWDII70SF - ARDPWT639037A, PWDII70SF - ARDPWT639038A, PWDII70SF - ARDPWT639039A, PWDII70SF - ARDPWT639040A, PWDII70DF - ARDPWT639042A, PWDII50DF - ARDPWT639043A, PWDII50DF - ARDPWT639044A, PWDII50DF - ARDPWT639045A, PWDII50DF - ARDPWT639046A, PWDII55DF - ARDPWT639048A, PWDII55DF - ARDPWT639049A, PWDII55DF - ARDPWT639050A, PWDII75DF - ARDPWT639052A, PWDII75DF - ARDPWT639053A, PWDII75DF - ARDPWT639054A, PWDII50SF - ARDPWT639056A, PWDII50SF - ARDPWT639057A, PWDII50SF - ARDPWT639058A, PWDII50SF - ARDPWT639059A, PWDII55DF - ARDPWT639061A, PWDII55DF - ARDPWT639063A, PWDII55SF - ARDPWT639064A, PWDII55SF - ARDPWT639066A, PWDII55SF - ARDPWT639067A, PWDII55SF - ARDPWT639068A, PWDII55SF - ARDPWT639069A, PWDII70DF - ARDPWT639071A, PWDII70DF - ARDPWT639072A, PWDII70DF - ARDPWT639073A, PWDII70DF - ARDPWT639074A, PWDII70DF - ARDPWT639076A, PWDII70SF - ARDPWT639077A, PWDII70SF - ARDPWT639078A, PWDII70SF - ARDPWT639079A, PWDII70SF - ARDPWT639080A, PWDII75DF - ARDPWT639082A, PWDII75DF - ARDPWT639084A, PWDII75SF - ARDPWT639086A, PWDII75SF - ARDPWT639088A, PWDII75SF - ARDPWT639089A, PWDII75SF - ARDPWT639090A, PWDII75SF - ARDPWT639091A, PWDII77DF - ARDPWT639093A, PWDII77DF - ARDPWT639095A, PWDII77DF - ARDPWT639096A, PWDII77DF - ARDPWT639098A, PWDII77DF - ARDPWT639099A, PWDII77SF - ARDPWT639101A, PWDII77SF - ARDPWT639103A, PWDII77SF - ARDPWT639104A, PWDII77SF - ARDPWT639105A, PWDII77SF - ARDPWT639106A, PWDII50DF - ARDPWT639108A, PWDII50DF - ARDPWT639109A, PWDII50DF - ARDPWT639110A, PWDII50DF - ARDPWT639111A, PWDII55DF - ARDPWT639113A, PWDII55DF - ARDPWT639114A, PWDII55DF - ARDPWT639115A, PWDII75DF - ARDPWT639117A, PWDII75DF - ARDPWT639118A, PWDII75DF - ARDPWT639119A, PWDII50SF -
Recall
- Recall Number
- Z-0757-2024
- Event Number
- 93567
- Firm
- Getinge Usa Sales Inc
- FEI Number
- 3013876692
- Product Code
- FTD
- Status
- Open, Classified
- Root Cause
- Equipment maintenance
- Initiated
- November 8, 2023
- Posted
- January 19, 2024
- Address
- 1 Geoffrey Way, Wayne, NJ, 07470-2035
Description
Maquet PowerLEDII OR Light System, Model Names and Catalog Numbers PWDII77SF - ARDPWT629100A, PWDII555DF - ARDPWT629101A, PWDII555SF - ARDPWT629102A, PWDII755DF - ARDPWT629103A, PWDII755SF - ARDPWT629104A, PWDII775DF - ARDPWT629105A, PWDII775SF - ARDPWT629106A, PWDII50DF - ARDPWT629109A, PWDII50SF - ARDPWT629110A, PWDII50SF - ARDPWT629111A, PWDII55DF - ARDPWT629112A, PWDII55SF - ARDPWT629113A, PWDII70DF - ARDPWT629114A, PWDII70DF - ARDPWT629115A, PWDII70SF - ARDPWT629116A, PWDII70SF - ARDPWT629117A, PWDII75DF - ARDPWT629118A, PWDII75SF - ARDPWT629119A, PWDII50DF - ARDPWT629120A, PWDII50SF - ARDPWT629121A, PWDII50SF - ARDPWT629122A, PWDII55DF - ARDPWT629123A, PWDII55SF - ARDPWT629124A, PWDII70DF - ARDPWT629125A, PWDII70DF - ARDPWT629126A, PWDII70SF - ARDPWT629127A, PWDII70SF - ARDPWT629128A, PWDII75DF - ARDPWT629129A, PWDII75SF - ARDPWT629130A, PWDII50DF - ARDPWT629131A, PWDII50SF - ARDPWT629132A, PWDII50SF - ARDPWT629133A, PWDII55DF - ARDPWT629134A, PWDII55SF - ARDPWT629135A, PWDII70DF - ARDPWT629136A, PWDII70DF - ARDPWT629137A, PWDII70SF - ARDPWT629138A, PWDII70SF - ARDPWT629139A, PWDII75DF - ARDPWT629140A, PWDII75SF - ARDPWT629141A, PWDII50DF - ARDPWT629142A, PWDII50SF - ARDPWT629143A, PWDII50SF - ARDPWT629144A, PWDII55DF - ARDPWT629145A, PWDII55SF - ARDPWT629146A, PWDII70DF - ARDPWT629147A, PWDII70DF - ARDPWT629148A, PWDII70SF - ARDPWT629149A, PWDII70SF - ARDPWT629150A, PWDII75DF - ARDPWT629151A, PWDII75SF - ARDPWT629152A, PWDII50DF - ARDPWT629154A, PWDII50DF - ARDPWT629155A, PWDII77SF - ARDPWT629156A, PWDII50DF - ARDPWT639001A, PWDII50DF - ARDPWT639002A, PWDII50DF - ARDPWT639003A, PWDII50DF - ARDPWT639004A, PWDII50SF - ARDPWT639006A, PWDII50SF - ARDPWT639007A, PWDII50SF - ARDPWT639008A, PWDII50SF - ARDPWT639009A, PWDII70DF - ARDPWT639011A, PWDII70DF - ARDPWT639012A, PWDII70DF - ARDPWT639013A, PWDII70DF - ARDPWT639014A, PWDII70SF - ARDPWT639016A, PWDII70SF - ARDPWT639017A, PWDII70SF - ARDPWT639018A, PWDII70SF - ARDPWT639019A, PWDII70DF - ARDPWT639021A, PWDII50DF - ARDPWT639022A, PWDII50DF - ARDPWT639023A, PWDII50DF - ARDPWT639024A, PWDII50DF - ARDPWT639025A, PWDII50SF - ARDPWT639027A, PWDII50SF - ARDPWT639028A, PWDII50SF - ARDPWT639029A, PWDII50SF - ARDPWT639030A, PWDII70DF - ARDPWT639032A, PWDII70DF - ARDPWT639033A, PWDII70DF - ARDPWT639034A, PWDII70DF - ARDPWT639035A, PWDII70SF - ARDPWT639037A, PWDII70SF - ARDPWT639038A, PWDII70SF - ARDPWT639039A, PWDII70SF - ARDPWT639040A, PWDII70DF - ARDPWT639042A, PWDII50DF - ARDPWT639043A, PWDII50DF - ARDPWT639044A, PWDII50DF - ARDPWT639045A, PWDII50DF - ARDPWT639046A, PWDII55DF - ARDPWT639048A, PWDII55DF - ARDPWT639049A, PWDII55DF - ARDPWT639050A, PWDII75DF - ARDPWT639052A, PWDII75DF - ARDPWT639053A, PWDII75DF - ARDPWT639054A, PWDII50SF - ARDPWT639056A, PWDII50SF - ARDPWT639057A, PWDII50SF - ARDPWT639058A, PWDII50SF - ARDPWT639059A, PWDII55DF - ARDPWT639061A, PWDII55DF - ARDPWT639063A, PWDII55SF - ARDPWT639064A, PWDII55SF - ARDPWT639066A, PWDII55SF - ARDPWT639067A, PWDII55SF - ARDPWT639068A, PWDII55SF - ARDPWT639069A, PWDII70DF - ARDPWT639071A, PWDII70DF - ARDPWT639072A, PWDII70DF - ARDPWT639073A, PWDII70DF - ARDPWT639074A, PWDII70DF - ARDPWT639076A, PWDII70SF - ARDPWT639077A, PWDII70SF - ARDPWT639078A, PWDII70SF - ARDPWT639079A, PWDII70SF - ARDPWT639080A, PWDII75DF - ARDPWT639082A, PWDII75DF - ARDPWT639084A, PWDII75SF - ARDPWT639086A, PWDII75SF - ARDPWT639088A, PWDII75SF - ARDPWT639089A, PWDII75SF - ARDPWT639090A, PWDII75SF - ARDPWT639091A, PWDII77DF - ARDPWT639093A, PWDII77DF - ARDPWT639095A, PWDII77DF - ARDPWT639096A, PWDII77DF - ARDPWT639098A, PWDII77DF - ARDPWT639099A, PWDII77SF - ARDPWT639101A, PWDII77SF - ARDPWT639103A, PWDII77SF - ARDPWT639104A, PWDII77SF - ARDPWT639105A, PWDII77SF - ARDPWT639106A, PWDII50DF - ARDPWT639108A, PWDII50DF - ARDPWT639109A, PWDII50DF - ARDPWT639110A, PWDII50DF - ARDPWT639111A, PWDII55DF - ARDPWT639113A, PWDII55DF - ARDPWT639114A, PWDII55DF - ARDPWT639115A, PWDII75DF - ARDPWT639117A, PWDII75DF - ARDPWT639118A, PWDII75DF - ARDPWT639119A, PWDII50SF -
A potential for a light system to fall in the operating room.
An URGENT MEDICAL DEVICE CORRECTION letter dated 11/8/23 was sent to customers. Actions to be taken by the customer: Our records indicate that you have been identified as having purchased at least one model of one of these product ranges. Please identify your affected units using the referenced devices on page 1 and follow the instructions below. Please distribute this notice and forward the link below to access manuals to anyone within your organization who may need to use the manuals, and especially to your service or service provider who performs your maintenance. Light system fixing and other replacements: Regarding: - the fixing of the light systems: suspensions fixing screws, adapter fixing screws, bushing fixing screws - the brake screws - the safety segments - the batteries, a dedicated instruction to illustrate and facilitate understanding of what to replace and what to order for replacement has been issued. The instruction is available on a dedicated space on the web portal https://www.getinge.com/int/campaigns/maintenance-and-service-on-or-light-system/, and provided upon request by emailing [email protected] Should you have questions or require additional information, please do not hesitate to contact Maquet SAS/Getinge or your local representative. Please complete and sign the attached MEDICAL DEVICE CORRECTION - RESPONSE FORM (page 7) to acknowledge that you have received this notification. Return the completed form to Maquet/Getinge by e-mailing a scanned copy to [email protected] or by faxing the form to (866) 350-0897. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. Actions to be taken by Maquet SAS / Getinge: As the majority of customers may not have regular contact with Getinge services or authorized Getinge representatives and therefore may not be aware of current maintenance recommendations, Maquet SAS/Getinge is providin
US Nationwide. Global Distribution.
236793 units in total