FDA Recall Terminated

API ZYM Bx2 (Ref 70493) Product Usage: ZYM B reagent is an additional test used for revealing the results of some miniaturized biochemical test including in the involved API strips.

Recall: Z-0756-2016 · Initiated January 8, 2016

Recall

Recall Number
Z-0756-2016
Event Number
73060
Firm
BioMerieux SA Chemin De L'Orme Marcy L'Etoile France
FEI Number
1000611339
Product Code
LIB
Status
Terminated
Root Cause
Other
Initiated
January 8, 2016
Posted
February 4, 2016
Terminated
July 14, 2017

Description

API ZYM Bx2 (Ref 70493) Product Usage: ZYM B reagent is an additional test used for revealing the results of some miniaturized biochemical test including in the involved API strips.

Reason

Discrepant identification results related to API identification strips. These discrepant identification results are due to false negative results on the tests of the strips linked to the ZYM B reagent.

Action

BioMrieux sent an Urgent Field Safety Notice letter DATED jANUARY 8, 2015 to affected customers. The letter identified the affected product, problem and action to be taken. dated . The letter indicated that they could continue to use product in inventory with mandatory requirements that before using a ZYM B ampule and each day of use of API strips that they perform a quality control with ATCC strains as described in each package insert of involved API strips. The letter also requested that they distribute this letter to all appropriate personnel in the laboratory, retain a copy for their files, and forward the information to all parties that may use the product. For additional assistance or have any questions, contact your local bioMrieux Customer Service representative.

Distribution

Nationwide Distribution in the states of: AL, CA, CO, CT, DC, FL, GA, HI, IL, IN, IA, KS, KY, ME, MD, MA, MI, MN, MS, MO, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TX, VT, VA, WA, WV, WI, WY

Quantity

13,271