ADVIA Centaur PSA Assay; Catalog Number: 06574155/SMN 10310292 (100 test kit) and Catalog Number: 02676506/SMN 10310293 (500 test kit) Product Usage: This in vitro diagnostic assay is intended to quantitatively measure prostate-specific Antigen (PSA) in human serum using the ADVIA Centaur/XP/XPT/CP Systems.

Recall: Z-0754-2016 · Initiated December 29, 2015

Recall

Recall Number
Z-0754-2016
Event Number
72968
Firm
Siemens Healthcare Diagnostics, Inc
FEI Number
1219913
Product Code
MTF
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
December 29, 2015
Posted
February 4, 2016
Terminated
November 29, 2019
Address
333 Coney St, East Walpole, MA, 02032-1516

Description

ADVIA Centaur PSA Assay; Catalog Number: 06574155/SMN 10310292 (100 test kit) and Catalog Number: 02676506/SMN 10310293 (500 test kit) Product Usage: This in vitro diagnostic assay is intended to quantitatively measure prostate-specific Antigen (PSA) in human serum using the ADVIA Centaur/XP/XPT/CP Systems.

Reason

PSA assay is not meeting the High Dose Hook Effect expectation in the Instructions for Use (IFU).

Action

An Urgent Medical Device Correction (UMDC) was sent to all affected Siemens Healthcare Diagnostics customers in the United States on December 30, 2015 to be delivered to customers on December 31, 2015. An Urgent Field Corrective Action (UFCA) was sent out for distribution to all customers outside the United States on December 29, 2015. These notices identified the affected product, problem and actions to be taken. For questions customers were instructed to contact Siemens Customer Care Center or local Siemens technical support representative.

Distribution

Worldwide - US Nationwide including Puerto Rico, OUS: Albania, Algeria, Angola, Armenia, Austria, Bahrain, Belarus, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Great Britain, Greece, Hungary, Iceland, Iraq, Italy, Jordan, Kazakhstan, Latvia, Lebanon, Lithuania, Luxembourg, Macedonia, Malta, Morocco, Netherlands, Norway, Oman, Pakistan, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Tunisia, Turkey, Turkmenistan, Unit.Arab Emir., Vatican, AR, AU, BD, BR, CA, CL, CN, CO, DO, EC, GT, HK, ID, IL, IN, JP, KR, LK, MM, MX, MY, NI, NP, NZ, PE, PH, PY, SA, SG, TH, TW, UY, VN and ZA.

Quantity

55,754 kits