FDA Recall Terminated

MCA5-07SD - 0.7MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochleostomy.

Recall: Z-0752-2014 · Initiated March 29, 2012

Recall

Recall Number
Z-0752-2014
Event Number
63624
Firm
The Anspach Effort, Inc.
FEI Number
1045834
Product Code
HBE
Status
Terminated
Root Cause
Labeling Change Control
Initiated
March 29, 2012
Posted
January 15, 2014
Terminated
March 11, 2015
Address
4500 Riverside Drive, Palm Beach Gardens, FL, 33410-4235

Description

MCA5-07SD - 0.7MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochleostomy.

Reason

The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide

Action

Firm is performing a Medical Device Recall by sending customers a recall letter and requesting the return of Anspach Micro Curved Attachments and Dissection Tools currently on hand.

Distribution

Worldwide Distribution: US (nationwide) and Internationally to: AU, BE, CA, CH, ES, FR, GB, IT, JP, NL, SE, and ZA.

Quantity

87