FDA Recall Open, Classified

Brand Name: Daytona Small Stature Product Name: CoCr LockingCap 4.5mm, Small Stature, T25, BT Model/Catalog Number: 79-0002 Software Version: N/A Product Description: Locking Cap Component: No

Recall: Z-0751-2025 · Initiated October 29, 2024

Recall

Recall Number
Z-0751-2025
Event Number
95760
Firm
SEASPINE ORTHOPEDICS CORPORATION
FEI Number
3012120772
Product Code
NKB
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
October 29, 2024
Posted
December 23, 2024
Address
5770 Armada Dr, Carlsbad, CA, 92008-4608

Description

Brand Name: Daytona Small Stature Product Name: CoCr LockingCap 4.5mm, Small Stature, T25, BT Model/Catalog Number: 79-0002 Software Version: N/A Product Description: Locking Cap Component: No

Reason

Due to manufacturing defects observed on Locking Caps that affect the ability for components to mate.

Action

Update 12/18/2024 - The firm sent an updated Customer Notification Letter on 12/12/2024 to customers that have not responded to the initial notification on 10/29/2024. On 10/29/2024, firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter to customer informing them that the firm has identified a nonconformance with the threads of the locking cap components that may be incongruous with optimal manufacturing specifications. The nonconformance can result in a locking cap that does not mate appropriately with the screw tulip. Customers are instructed to: -Review their inventory and return the affected part(s) using the provided label to return the inventory in their possession. Any questions contact Dalton Daggs at [email protected] or Danielle Nelson-Archer at [email protected].

Distribution

U.S. Nationwide distribution in the states of CA, CO, FL, GA, PA, TX, and VA.

Quantity

92 devices