FDA Recall Terminated

Mobilett XP, mobile X-ray system, Model number 1818454

Recall: Z-0749-2007 · Initiated March 21, 2007

Recall

Recall Number
Z-0749-2007
Event Number
37698
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
IZL
Status
Terminated
Root Cause
Other
Initiated
March 21, 2007
Posted
April 19, 2007
Terminated
August 7, 2007
Address
51 Valley Stream Pkwy, Malvern, PA, 19355-1406

Description

Mobilett XP, mobile X-ray system, Model number 1818454

Reason

Mounting bolts for tank fork assembly and collimator flange may become loose.

Action

Siemens Service Personnel are visiting each consignee, beginning March 21, 2007, to replace the required mounting bolts and tighten them according to required torque specifications.

Distribution

The products were shipped to medical facilities nationwide.

Quantity

186 units