FDA Recall Terminated

Olympus Electrosurgical Unit. The Olympus Electrosurgical Unit (ESG-100) is used in conjunction with cutting and coagulation electrosurgical devices intended for use in electrosurgery procedures during flexible gastrointestinal endoscopy.

Recall: Z-0747-2012 · Initiated November 28, 2011

Recall

Recall Number
Z-0747-2012
Event Number
60548
Firm
Olympus America Inc.
FEI Number
2429304
Product Code
GEI
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
November 28, 2011
Posted
January 13, 2012
Terminated
March 1, 2013
Address
3500 Corporate Pkwy. P.O. Box 610, Center Valley, PA, 18034-0610

Description

Olympus Electrosurgical Unit. The Olympus Electrosurgical Unit (ESG-100) is used in conjunction with cutting and coagulation electrosurgical devices intended for use in electrosurgery procedures during flexible gastrointestinal endoscopy.

Reason

Investigations of customer complaints revealed a counterfeit capacitor was used in the manufacture of some lots of the ESG-100.

Action

Olympus America Inc. sent an URGENT: MEDICAL DEVICE CORRECTION ACTION" letter dated November 28, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to discontinue use of the affected product. A questionnaire was attached to the letter for customers to complete and return via fax at 484-896-7128. Contact the firm at 484-896-5000 for questions regarding this notice.

Distribution

Nationwide Distribution.

Quantity

245