FDA Recall Terminated

Harmony Insight Diagnostic Monitor Support System - 100 Spring Arm (EMS) The system is intended for use in surgical operating rooms (OR) or intraoperative surgical/imaging rooms where simultaneous multiple image viewing is required

Recall: Z-0746-2017 · Initiated October 5, 2016

Recall

Recall Number
Z-0746-2017
Event Number
75385
Firm
Steris Corporation
FEI Number
1527821
Product Code
FTD
Status
Terminated
Root Cause
Component design/selection
Initiated
October 5, 2016
Posted
December 12, 2016
Terminated
April 14, 2017
Address
5960 Heisley Rd, Mentor, OH, 44060-1834

Description

Harmony Insight Diagnostic Monitor Support System - 100 Spring Arm (EMS) The system is intended for use in surgical operating rooms (OR) or intraoperative surgical/imaging rooms where simultaneous multiple image viewing is required

Reason

The Drager Pendula spring arm device has a set of screws located on the spring arm that were not properly secured with Loctite and may become loose.

Action

Steris sent an Urgent Voluntary Field Correction Notice to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to remind staff personnel that Harmony Insight Diagnostic Monitor Support Systems, including the affected Drager Pendula spring arms, are not designed or intended to be placed over hospital personnel or patients and should be moved with care. Customers with questions were instructed to call 440-392-7358 or Steris Customer Service at 1-800-548-4873. For questions regarding this recall call 440-392-7231.

Distribution

Nationwide Distribution including AK, AZ, CA, CO, FL, IL, KY, MI, MO, NB, NJ, NY, OH, PA, RI, SD, TN, TX, UT, WA & WV.

Quantity

63