FDA Recall Open, Classified

TEMA Elbow - Axle Small REF 1590.15.010 TEMA Elbow - Axle Large REF 1590.15.020 TEMA Elbow System is a cemented total elbow prosthesis.

Recall: Z-0744-2025 · Initiated October 2, 2024

Recall

Recall Number
Z-0744-2025
Event Number
95732
Firm
Limacorporate S.p.A Via Nazionale 52 Di San Daniele San Daniele Del Friuli Italy
FEI Number
3008021110
Product Code
JDB
Status
Open, Classified
Root Cause
Process control
Initiated
October 2, 2024
Posted
December 20, 2024

Description

TEMA Elbow - Axle Small REF 1590.15.010 TEMA Elbow - Axle Large REF 1590.15.020 TEMA Elbow System is a cemented total elbow prosthesis.

Reason

Due to a potential difference in mechanical behavior between the implanted device and tool(s) used during revision surgery.

Action

On 10/02/2024, the firm notified U.S. surgeons of the recall and the two U.S. distributors were informed via email on 10/11/2024. An updated/follow-up Recall Notification and Acknowledgement form is pending. The update/follow-up Recall Notification is informing surgeons and distributors that the firm has received a total of two similar complaints, regarding issues related to difficulty in disassembly of the axle component by using the screwdriver as per Surgical Technique. According to the internal investigation, the main concurring root causes of the problem seem to be the incomplete shot peening on the taper and the missing identification of a critical dimension (thread length). The Recall Notification informs surgeons to use other instruments available in the Operating Room to successfully complete the surgery or, in the worst case, several parts/the entire implant (both the humeral and the ulnar components) should be removed. Surgeons are also asked to Fill out, sign, and send the following Acknowledgement and Receipt Form via email as confirmation that You have read and acknowledged the content of this document. For questions or assistance, contact: Contact of the Importer: Kenneth Newman - Senior Regulatory Affairs Specialist (LimaCorporate now Enovis) /Recall Coordinator (Enovis) E-mail: [email protected] Or to the Manufacturer contact point: Federica Malvaso Post-Market Surveillance Lead (LimaCorporate now Enovis) E-mail: [email protected] Distributors are instructed to: Check your stock to locate and quarantine the affected devices. Devices must be sent back to: Lima U.S.A., Inc. | 2001 NE Green Oaks Blvd. Ste.100 | Arlington, Texas 76006 | USA within 15 days together with a signed hard copy of the following Acknowledgement and Receipt Form.

Distribution

U.S: FL and TX O.U.S.: N/A

Quantity

~87 devices (U.S.)