FDA Recall Terminated

Vis-U-All High Temperature Tubing 4"x100'. One roll per box; 10 boxes per case 10 rolls/cs Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is designed to maintain sterility of properly processed medical devices during normal handling and storage until the pouch is opened and the medical device is removed for use.

Recall: Z-0744-2018 · Initiated June 15, 2017

Recall

Recall Number
Z-0744-2018
Event Number
79100
Firm
Steris Corporation
FEI Number
1527821
Product Code
FRG
Status
Terminated
Root Cause
Component change control
Initiated
June 15, 2017
Terminated
July 17, 2018
Address
5960 Heisley Rd, Mentor, OH, 44060-1834

Description

Vis-U-All High Temperature Tubing 4"x100'. One roll per box; 10 boxes per case 10 rolls/cs Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is designed to maintain sterility of properly processed medical devices during normal handling and storage until the pouch is opened and the medical device is removed for use.

Reason

On June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.

Action

Steris sent an Urgent Voluntary Recall Notice letter dated July 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately inspect on-hand inventory for recalled product. If you have product in your possession, please quarantine it and contact firm to coordinate shipment of replacement product. If you have questions regarding this matter, please contact STERIS Customer Service at 1-800-548-4873.

Distribution

US Nationwide Distribution

Quantity

51