FDA Recall Terminated

PTS Panels CHOL+HDL+GLU Test Strips, Reference Number 5412, for use on CardioChek PA or CardioChek Plus analyzers

Recall: Z-0741-2020 · Initiated March 15, 2016

Recall

Recall Number
Z-0741-2020
Event Number
84264
Firm
Polymer Technology Systems, Inc.
FEI Number
3015525148
Product Code
NBW
Status
Terminated
Root Cause
Storage
Initiated
March 15, 2016
Terminated
June 5, 2020
Address
7736 Zionsville Rd, Indianapolis, IN, 46268-2175

Description

PTS Panels CHOL+HDL+GLU Test Strips, Reference Number 5412, for use on CardioChek PA or CardioChek Plus analyzers

Reason

This lot of test strips is showing an under-recovery when tested against a reference method.

Action

Customer Notification Letters dated 3/15/16 was sent to customers. Distributors were instructed to do the following: We need your assistance to communicate this information to any customers to whom you have shipped this specific lot of product. As per Quality System Regulations, you must have full traceability of medical product to the end-user. Please advise your customers to complete the Customer Acknowledgement and Notification Form so that PTS Diagnostics can document that they have received the notification. Upon receipt of the acknowledgement, PTS Diagnostics will provide replacements for any product which has not been consumed. Customers with any of this product remaining in their possession must Immediately discontinue use and dispose of their remaining inventory of PTS Panels CHOL+HOL+GLU test strips, lot number 1502. No other lots of this product or other PTS Diagnostics products are impacted by this action. PTS Diagnostics will provide appropriate replacement product for any product properly identified on the attached Customer Acknowledgement and Confirmation Form. Please immediately identify the number of remaining PTS Panels CHOL+HDL+GLU test strip vials you currently have remaining in inventory of lot 1502 on the attached form. We will supply you with replacement product for any unused inventory you have of this product and lot. Please return your completed form to PTS Diagnostics via fax at +1-317-870-5608 or via email at [email protected]. Customers were instructed to do the following: Customers with any of this product remaining in their possession must Immediately discontinue use and dispose of their remaining inventory of PTS Panels CHOL+HOL+GLU test strips, lot number 1502. No other lots of this product or other PTS Diagnostics products are impacted by this action. PTS Diagnostics will provide appropriate replacement product for any product properly identified on the attached Customer Acknowledgement and Confir

Distribution

The products were distributed to the following US states: CA, CT, FL, IN, MA, MD, MI, MN, NC, NJ, NM, NY, OR, PA, TN, and WI. The products were distributed to the following foreign countries: Australia, Belgium, Indonesia, Italy, Poland, South Africa, UK, and Vietnam.

Quantity

6467