FDA Recall Terminated

HAMILTON-T1 Ventilator with software versions 1.1.2 or lower. The HAMILTON T-1 Ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics

Recall: Z-0741-2013 · Initiated December 19, 2012

Recall

Recall Number
Z-0741-2013
Event Number
64117
Firm
Hamilton Medical Inc
FEI Number
2937708
Product Code
CBK
Status
Terminated
Root Cause
Software design
Initiated
December 19, 2012
Posted
January 31, 2013
Terminated
April 22, 2014
Address
4990 Energy Way P O Box 30008, Reno, NV, 89502-4123

Description

HAMILTON-T1 Ventilator with software versions 1.1.2 or lower. The HAMILTON T-1 Ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics

Reason

Analysis of a customer complaint has shown that during ventilation of small pediatric patients with high airway resistance and low lung compliance, the oxygen consumption of a Hamilton T1 ventilator with software versions 1.1.2 and lower must be calculated using a larger margin than originally expected.

Action

Hamilton sent a Field Safety alert letter dated January 11, 2013 to all customers. The letter identified the affected product, explained the potential problem and a included Medical Device Safety Alert and Corrective Action document. Hamilton is working on a new software version for this issue. Customers are asked to update their operator manual with the update oxygen consumption formula page which provided revised information for calculating oxygen capacity. When the new software is available, Hamilton will provide loaner devices to each location while their unit is being upgraded. Questions should be addressed to Hamilton Medical at 800-426-6331 ext 215

Distribution

US Nationwide Distribution including the country of Switzerland.

Quantity

97 total ventilators; 69 of part number 161005 and 28 of part number 161006