FDA Recall Terminated

Urine Electrolyte Diluent, Product Code: 1112352 - Product Usage: used to dilute urine specimens for analysis of sodium (Na+) and potassium (K+) on VITROS 250/350/5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems.

Recall: Z-0740-2020 · Initiated October 14, 2019

Recall

Recall Number
Z-0740-2020
Event Number
84123
Firm
Ortho Clinical Diagnostics Inc
FEI Number
2250051
Product Code
KSG
Status
Terminated
Root Cause
Device Design
Initiated
October 14, 2019
Terminated
November 13, 2020
Address
1001 US Highway 202, Raritan, NJ, 08869-1424

Description

Urine Electrolyte Diluent, Product Code: 1112352 - Product Usage: used to dilute urine specimens for analysis of sodium (Na+) and potassium (K+) on VITROS 250/350/5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems.

Reason

Potential for the instability of the diluents VITROS 7% BSA (Product Code 8262487) and VITROS Urine Electrolyte Diluent (UED) (Product Code 1110352) when stored in a 2mL cup for 7 days.

Action

On August 9, 2019 the firm distributed Urgent Medical Device Recall Communication letters by via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) summarizing the reason as Ortho Clinical Diagnostics determined that VITROS 7% BSA and VITROS Urine Electrolyte Diluent (UED) do not meet the current OAS time of d 7 days when stored on the VITROS 250/350 System in a 2mL cup only. Orthos internal testing compared results from samples diluted with diluent stored on the analyzer versus fresh diluent. When using 7% BSA or UED stored in a 2mL cup for 7 days, the differences in the results vs fresh diluent were >10% for some assays. It explained as a resolution: " The OAS of 7% BSA and UED in 2mL cups has been revised to d24 hours. " Current OAS time remains at d7 days for storage in 5mL and 10 mL bottles. " The Instructions for Use and appropriate User Documentation will be revised at a future date. It instructs them to: - Send the enclosed customer letter to all customers to whom you shipped or know to have a VITROS 250/350 System. Please continue to provide this letter to new customers until the Instructions for Use are updated. - Complete and return the Confirmation of Receipt form by October 22, 2019.

Distribution

Worldwide distribution - US Nationwide distribution and countries BM, AU, BR, CA, CL, CN, CO, IN, JP, MX, SG, GB, FR, DE, IT, ES, PT, PL, RU, DK, NO, SE, BE, NL.

Quantity

2,573