FDA Recall Terminated

NEXES VEN IVIEW DAB DET KT JPN-US EXPORT, Ventana Part Number 760-041, Roche GMMI 05266084001 in vitro diagnostic

Recall: Z-0740-2018 · Initiated December 19, 2017

Recall

Recall Number
Z-0740-2018
Event Number
78942
Firm
Ventana Medical Systems Inc
FEI Number
1000125183
Product Code
NJT
Status
Terminated
Root Cause
Process control
Initiated
December 19, 2017
Posted
February 27, 2018
Terminated
November 22, 2021
Address
1910 E Innovation Park Dr, Oro Valley, AZ, 85755-1962

Description

NEXES VEN IVIEW DAB DET KT JPN-US EXPORT, Ventana Part Number 760-041, Roche GMMI 05266084001 in vitro diagnostic

Reason

Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.

Action

On Dec 19th all Roche affiliates (the direct consignees) worldwide were notified via Safety Board Notification (SBN) and subsequent Field Safety Notification (EX-US) and UMDC TD00090 (US) SBN and FSN revised Dec 20th due to 3 affected part numbers missed on original letter. The instructions are as follows: "Actions taken by Roche Diagnostics All affected product has been placed on hold. Ventana has reworked all product in its inventory and is in the process of manufacturing new lots for distribution and replacement of customer affected kits. Customers will be notified when corrected product is available. Actions to be taken by the customer/user Affected kits may continue to be used by customers until corrected product is available, however Ventana is mandating that the affected IHC detection kits (iView, ultraView, OptiView) must only be used in conjunction with same-slide controls. These controls must be appropriate for each assay and capable of detecting false negative results due to a complete or partial reagent dispense failure. CINtec PLUS Cytology does not have the capacity for same slide controls, so system-level controls should be maintained. ultraView SISH Detection is used for HER2 analysis, and employs internal positive controls; no external control is required. For assays that directly relate to clinical therapy decision making (e.g. ER/PR, HER2, ALK, etc.), it is additionally important to select a same slide positive control tissue with sufficient sensitivity to detect small decreases in intensity that may cause borderline positive cases to appear as negative (e.g. HER2 2+ vs. 1+). Although the use of same slide controls is considered optimal laboratory practice and strongly recommended by Ventana, customers may revert to standard run controls once non-impacted product is received. In order to reduce the risk of this issue impacting patient care, customers not using same slide controls as a standard practice should follow their

Distribution

Worldwide distribution, including US nationwide, Argentina, Australia, Hong Kong, Indonesia, Philippines, Brazil, Chile, Thailand, Korea, Malaysia, New Zealand, Ecuador, Mexico, Pakistan, Vietnam, Germany, Japan, India, Colombia, Singapore, Uruguay, Peru, Canada, Taiwan, China,

Quantity

3122 units