FDA Recall Terminated

ABX PENTRA Glucose PAP CP ref. A11A01668 reagent is intended for the quantitative in vitro diagnostic determination of glucose in human serum, plasma and urine using glucose oxidase method by colorimetry.

Recall: Z-0740-2016 · Initiated January 11, 2016

Recall

Recall Number
Z-0740-2016
Event Number
73033
Firm
Horiba Instruments Inc
FEI Number
3002698356
Product Code
CFR
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
January 11, 2016
Posted
February 2, 2016
Terminated
June 7, 2016
Address
9755 Research Dr, Irvine, CA, 92618-4626

Description

ABX PENTRA Glucose PAP CP ref. A11A01668 reagent is intended for the quantitative in vitro diagnostic determination of glucose in human serum, plasma and urine using glucose oxidase method by colorimetry.

Reason

N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.

Action

Customer notification letters were sent on 01/11/16. Customers were provided the following instructions/actions: When performing tests that use the Tinder reaction, the blood draw should be done before treatment is given to the patient. Reagent inserts for the concerned HORIBA Medical Products have been updated and now include the following additional warning: Patients treated with N-Acetylcysteine (NAC) for Paracetamol overdose may generate a false low result. The letter request customers share the information with laboratory staff and retain the notification as part of their laboratory documentation. The letter states it is mandatory for customers to complete and return the enclosed response form within 10 days in order for the firm to verify the understanding of the notice. If any questions contact the local HORIBA Medical representative.

Distribution

U.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.

Quantity

216