FDA Recall Terminated

Philips Collimators, Model numbers: 9896 010 00614, 9896 010 00615, 9804 607 00301, and 9804 602 83501. Product is a Collimator/ Beam Limiting Device and are replacement components for the Bucky Diagnost TH X-ray system.

Recall: Z-0739-2007 · Initiated March 15, 2007

Recall

Recall Number
Z-0739-2007
Event Number
37767
Firm
Philips Medical Systems North America Co. Phillips
FEI Number
1000524572
Product Code
IZW
Status
Terminated
Root Cause
Other
Initiated
March 15, 2007
Posted
June 14, 2007
Terminated
November 15, 2010
Address
22100 Bothell Everett Hwy, Bothell, WA, 98021-8431

Description

Philips Collimators, Model numbers: 9896 010 00614, 9896 010 00615, 9804 607 00301, and 9804 602 83501. Product is a Collimator/ Beam Limiting Device and are replacement components for the Bucky Diagnost TH X-ray system.

Reason

The Collimator field-defining light beam does not meet the light field contrast ratio requirement of 4:1. System is non-compliant with performance standard (1020.31 (d)(2)(iii) 21 CFR

Action

Firm issued letter on 03/20/2007 to consignees advising them that firm representatives will contact consignee to replace collimators at no charge.

Distribution

Nationwide; Units were distributed to hospitals, radiologists, and health care clinics in IA, LA, ND, PA, and TX, and Washington DC.

Quantity

6